FDA Medical Device Recalls (openFDA)

Smiths Medical ASD Inc.: Tube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck Flange 4.5mm - Product Usage: intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed for single-patient use up to 5 times.

Recall number
Z-1145-2021
Recalling firm
Smiths Medical ASD Inc.
Product
Tube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck Flange 4.5mm - Product Usage: intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed for single-patient use up to 5 times.
Status
Terminated
Initiated
2021-01-18
Posted
not on file
Terminated
2022-12-20
Reason
Smiths Medical became aware that three lot numbers of a specific model of Bivona Cuffless FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) may have been incorrectly packaged with a straight neck flange instead of a V-neck flange. As a result, there is potential for patient injury if an incorrect device or accessory is selected for use. This use could lead to an internal injury resulting from excessive device contact with tissue, or an external patient injury due to external portions of the device pressing into or rubbing against tissue.
Root cause
Under Investigation by firm
Action
Distribution of the "URGENT MEDICAL DEVICE RECALL NOTICE" dated 01/29/2021 during the week of January 18, 2021.
Quantity
548 units
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, IL, KS, KY, LA, MA, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, TX, UT, VA, and WA.
Lot, serial or model codes
model 60PFS45, lots 3991171 (202 units), 3991172 (117 units), and 3988103 (229 units).
510(k) numbers
["K944178"]
PMA numbers
not on file
Product code
JOH
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register