Integra LifeSciences Corp.: Tibiaxys Ankle Arthrodesis Plate Compression Guide, Model #159635, part # 159 635, Batch #: EAHZ, EAJ1, ELGV, ENQE & ENQH. Product Usage: The Compression guide is a re-usable instrument used in the implantation of the Tibiaxys product. The Tibiaxys system is intended for fixation of bone fractures or for bone reconstruction (arthrodesis, osteotomies, and fractures of ankle joint, distal tibia and fibula).
- Recall number
- Z-1144-2012
- Recalling firm
- Integra LifeSciences Corp.
- Product
- Tibiaxys Ankle Arthrodesis Plate Compression Guide, Model #159635, part # 159 635, Batch #: EAHZ, EAJ1, ELGV, ENQE & ENQH. Product Usage: The Compression guide is a re-usable instrument used in the implantation of the Tibiaxys product. The Tibiaxys system is intended for fixation of bone fractures or for bone reconstruction (arthrodesis, osteotomies, and fractures of ankle joint, distal tibia and fibula).
- Status
- Terminated
- Initiated
- 2011-11-17
- Posted
- 2012-03-02
- Terminated
- 2013-02-19
- Reason
- During 6/30/2010 to 8/17/2011 the firm received five (5) incident complaints of breakage of the threaded part of the compression guide which potentially occurred when: 1) a compression is applied between talus and tibia. (Step 3 of the product description) or when 2) the surgeon screws the compression guide into the plate which most likely was attributed to improper surgical techniques used durin
- Root cause
- Nonconforming Material/Component
- Action
- Integra sent a "Field Safety Notice" letter dated November 17, 2011 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review the Surgical Techinique Manual and follow the recommended guiidelines to avoid breakage. ensure that this information is distributed to their surgical teams, and fill out the enclosed "Field Safety Notice Acknowledgement and Return Form" and return to bonnie.kuzbyt@integralife.com or fax to 1-609-750-7999. For questions call 609-936-6932 or email at sean.luland@integralife.com.
- Quantity
- 168 units
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of; AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KY, MA, MD, MI, MO, MN, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, WA & WI. Product was also shipped to the following countries: Austria, Belgium, Canada, Czech Republic, France, Germany, Ireland, Israel, Italy, Lebanon, Saudi Arabia, South Africa, Spain, Sweden, Switzerland & United Kingdom.
- Lot, serial or model codes
- Model #159635, part # 159 635, Batch #: EAHZ, EAJ1, ELGV, ENQE & ENQH.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OYK
- Firm FEI
- 3003418325
- Firm city
- Plainsboro
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28