FDA Medical Device Recalls (openFDA)
C.R. Bard, Inc., Urological Division: Bardex Foley Catheter, 22 Fr 40cc- 20cc Ribbed Balloons, Coleman Model Red Latex, 3 Way Radiopaque, Lubricath Coated, Sterile, Single Use Device, C.R. Bard, Inc., Covington, GA 30014. The catheter is an open whistle tip with two eyes on the same side. It is made of red latex for added catheter stiffness and radiopacity, with a Lubricath hydrophilic coating. This product has 3 lumens; a lumen for drainage, a lumen for inflation of the bladder balloon, and a lumen for inflation of the prostatic balloon.
- Recall number
- Z-1144-03
- Recalling firm
- C.R. Bard, Inc., Urological Division
- Product
- Bardex Foley Catheter, 22 Fr 40cc- 20cc Ribbed Balloons, Coleman Model Red Latex, 3 Way Radiopaque, Lubricath Coated, Sterile, Single Use Device, C.R. Bard, Inc., Covington, GA 30014. The catheter is an open whistle tip with two eyes on the same side. It is made of red latex for added catheter stiffness and radiopacity, with a Lubricath hydrophilic coating. This product has 3 lumens; a lumen for drainage, a lumen for inflation of the bladder balloon, and a lumen for inflation of the prostatic balloon.
- Status
- Terminated
- Initiated
- 2002-08-29
- Posted
- 2003-08-21
- Terminated
- 2003-09-19
- Reason
- On some units the inflation valve labeling is reversed on the Prostate and Bladder balloon inflation ports.
- Root cause
- Other
- Action
- Consignees were notified by letter on 8/29/2002. They were advised to pretest the catheter prior to inserting the catheter into the urethra.If the pretest indicates that the catheter balloons do not inflate as labeled, they were advised not to use the product and to contact Bard Medical Division Field Assurance Manager to have the product returned and credited.
- Quantity
- 1461 units
- Distribution
- The product was distributed nationwide, Canada, Belgium and Germany.
- Lot, serial or model codes
- Catalog Number 6016L22 (all units packaged from January 1999 and shipped prior to September 2002).
- 510(k) numbers
- ["K984136"]
- PMA numbers
- not on file
- Product code
- MJC
- Firm FEI
- 1018233
- Firm city
- Covington
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28