Skip to the content

FDA Medical Device Recalls (openFDA)

C.R. Bard, Inc., Urological Division: Bardex Foley Catheter, 22 Fr 40cc- 20cc Ribbed Balloons, Coleman Model Red Latex, 3 Way Radiopaque, Lubricath Coated, Sterile, Single Use Device, C.R. Bard, Inc., Covington, GA 30014. The catheter is an open whistle tip with two eyes on the same side. It is made of red latex for added catheter stiffness and radiopacity, with a Lubricath hydrophilic coating. This product has 3 lumens; a lumen for drainage, a lumen for inflation of the bladder balloon, and a lumen for inflation of the prostatic balloon.

Recall number
Z-1144-03
Recalling firm
C.R. Bard, Inc., Urological Division
Product
Bardex Foley Catheter, 22 Fr 40cc- 20cc Ribbed Balloons, Coleman Model Red Latex, 3 Way Radiopaque, Lubricath Coated, Sterile, Single Use Device, C.R. Bard, Inc., Covington, GA 30014. The catheter is an open whistle tip with two eyes on the same side. It is made of red latex for added catheter stiffness and radiopacity, with a Lubricath hydrophilic coating. This product has 3 lumens; a lumen for drainage, a lumen for inflation of the bladder balloon, and a lumen for inflation of the prostatic balloon.
Status
Terminated
Initiated
2002-08-29
Posted
2003-08-21
Terminated
2003-09-19
Reason
On some units the inflation valve labeling is reversed on the Prostate and Bladder balloon inflation ports.
Root cause
Other
Action
Consignees were notified by letter on 8/29/2002. They were advised to pretest the catheter prior to inserting the catheter into the urethra.If the pretest indicates that the catheter balloons do not inflate as labeled, they were advised not to use the product and to contact Bard Medical Division Field Assurance Manager to have the product returned and credited.
Quantity
1461 units
Distribution
The product was distributed nationwide, Canada, Belgium and Germany.
Lot, serial or model codes
Catalog Number 6016L22 (all units packaged from January 1999 and shipped prior to September 2002).
510(k) numbers
["K984136"]
PMA numbers
not on file
Product code
MJC
Firm FEI
1018233
Firm city
Covington
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register