Voco GmbH: Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement
- Recall number
- Z-1143-2024
- Recalling firm
- Voco GmbH
- Product
- Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement
- Status
- Open, Classified
- Initiated
- 2024-01-25
- Posted
- 2024-02-16
- Terminated
- not on file
- Reason
- Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.
- Root cause
- Process control
- Action
- Voco issued a field safety notice to its consignees on 01/25/2024 via email. The notice explained the problem, risk, and requested the product be returned. Those who have sold the product were directed to notify their customers and collect the affected product. For questions contact Mr. Michael Cyll at m.cyll@voco.de.
- Quantity
- 1021 units
- Distribution
- Worldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS (foreign) to countries of: Belgium, Bulgaria, Denmark, Germany, France, Greece, Great Britain, India, Italy, Canada, Quatar, Lithuania, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic, Australia, Saudi Arabia, & Ireland.
- Lot, serial or model codes
- a) 2115, UDI/DI +E22121151, Lot Numbers: 2346779, 2401515; b) 2117, UDI/DI +E22121171, Lot Numbers: 2347832; c) 2118, UDI/DI +E22121181, Lot Numbers: 2349536; d) 2119, UDI/DI +E22121191, Lot Numbers: 2348501; e) 2120, UDI/DI +E22121201, Lot Numbers: 2350638; f) 2122, UDI/DI +E22121221, Lot Numbers: 2349527
- 510(k) numbers
- ["K090817"]
- PMA numbers
- not on file
- Product code
- EMA
- Firm FEI
- 3002718942
- Firm city
- Cuxhaven
- Firm state
- not on file
- Firm country
- Germany
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28