Skip to the content

FDA Medical Device Recalls (openFDA)

FHC, Inc.: FHC microTargeting Guideline 4000 ver 1.4 (GL4K) Software, Catalog # MT-GL4K. Intended to assist in functional neurological procedures.

Recall number
Z-1143-2010
Recalling firm
FHC, Inc.
Product
FHC microTargeting Guideline 4000 ver 1.4 (GL4K) Software, Catalog # MT-GL4K. Intended to assist in functional neurological procedures.
Status
Terminated
Initiated
2009-12-28
Posted
2010-03-16
Terminated
2010-03-18
Reason
The raster tab for the GL4K version 1.4 software may assign incorrect depths for waveform recordings and if used to identify the implantation target for the DBS lead, the implant may be placed in the wrong location.
Root cause
Software change control
Action
FHC notified customers by telephone and by an Urgent: Medical Device Correction email, dated December 2009, beginning December 28th. Customers have been instructed to keep the instruction with the GL4K until FHC can provide a revision to the software; and, to use only "distance from zero" or "distance traveled" modes (not "distance from target" mode) during procedures using the raster tab. Customers are to notify their FHC representative to confirm receipt of the notification. A software patch has been verified and validated and will be installed on all affected units. The new version of the corrected software is Guideline 4000 version 1.41. Questions should be directed to Jason Moyer, product manager or Keri Seitz, recall coordinator at 207-666-8190.
Quantity
10
Distribution
Worldwide Distribution -- USA, including states of GA, MA, NC, NJ, OH, RI, TN, TX, and WV and Bucharest, Romania.
Lot, serial or model codes
Software ver 1.4, Serial Numbers: 19454, 14423, 17349, 19215, 15573, 19424, 17583, 17589, 17588, and 17581. This version of software was released in August 2009
510(k) numbers
["K071364"]
PMA numbers
not on file
Product code
GZL
Firm FEI
3002250546
Firm city
Bowdoinham
Firm state
ME
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register