Skip to the content

FDA Medical Device Recalls (openFDA)

HF Acquisition Co LLC: Biological, Bio52 HealthFirst Mail In Monitoring System, 52 Tests, Item: 1003600. Biological, Bio12 HealthFirst Mail In Monitoring System, 12 Tests, Item: 1003580

Recall number
Z-1141-2025
Recalling firm
HF Acquisition Co LLC
Product
Biological, Bio52 HealthFirst Mail In Monitoring System, 52 Tests, Item: 1003600. Biological, Bio12 HealthFirst Mail In Monitoring System, 12 Tests, Item: 1003580
Status
Open, Classified
Initiated
2024-12-11
Posted
2025-02-11
Terminated
not on file
Reason
Spore test strip monitoring system, designed for routine monitoring of sterilization processes, may have a printing error on the return envelope that may cause spore test results to be reported to the incorrect customer's account.
Root cause
Error in labeling
Action
On 12/11/2024, correction notices were mailed to customers asking them to do the following: 1. Check the box number on your Biological Weekly Monitoring kit for the affected lot numbers. The box number is located under the bar code on the outside of the box (see image below). 3. Check the serial number on all test strip return envelopes located in the box. This number is the last set of digits in the alphanumeric string located to the left of the bar code on the envelope. 4. Verify that the box number matches the last three digits of the serial number on the return envelope. 5. If the numbers do not match, you are hereby instructed to isolate your inventory of the above-listed lot numbers and dispose of it in regular trash. Firm will work with you on replacement product. 6. Ensure this letter is promptly brought to the attention of those responsible for managing the medical device inventory. Distribute this notification to all device users within your facility and network to ensure they are informed of this field correction letter. Please share this notification with any organizations that have received the potentially affected devices. 7. Complete and return the response form via email to quality@healthfirst.com Should you have any questions about this urgent field correction, please contact Customer Service at 855-476-1342 or email us at customerservice@healthfirst.com.
Quantity
4,021
Distribution
US: WI, FL, NY, CO, NJ, OH, IN, CA, WA, TX, AZ, MD, MN, NC, MI, MS, ID, TN, KS, SC, RI, LA, GA, PA, NH, UT, VA, MA, NV, MO, IL, SD, AR, OR, HI, NE, DC, CT, NM, ND, MT, OK, AL, VT, WY, KY, ME, IA, WV, AK
Lot, serial or model codes
Item/UDI-DI/Lot(Expiration): 1003600/10850065007820/ 2301060(11/30/2024) 2302261(11/30/2024) 2307261(12/31/2024 2312260(1/31/2025) 2313062(1/31/2025) 2331880(6/30/2025) 2401664(8/31/2025) 2414060(2/28/2026) 2414862(3/31/2026) 2414870(3/31/2026) 2416470(3/31/2026) 2416481(4/30/2026) 2416482(4/30/2026) 1003580/10850065007813/2421181(5/31/2026)
510(k) numbers
["K041099","K905425","K023716"]
PMA numbers
not on file
Product code
FRC
Firm FEI
3014534
Firm city
Mukilteo
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register