FDA Medical Device Recalls (openFDA)
Ecolab Inc: Slush + Warmer Disc Drape; Model ORS-320 Slush + Warmer Disc Drapes - Product Usage: single-use equipment cover intended for use during various surgeries where warm irrigation, slush and/or cold solution is required.
- Recall number
- Z-1141-2021
- Recalling firm
- Ecolab Inc
- Product
- Slush + Warmer Disc Drape; Model ORS-320 Slush + Warmer Disc Drapes - Product Usage: single-use equipment cover intended for use during various surgeries where warm irrigation, slush and/or cold solution is required.
- Status
- Terminated
- Initiated
- 2021-01-11
- Posted
- not on file
- Terminated
- 2023-01-09
- Reason
- Slits were identified in a low quantity of ORS-320 Slush + Warmer Disc Drapes from a single lot. A source for potential manufacturing error was identified and investigation is underway to prevent further recurrence. The slits may lead to a breach in the sterile barrier between fluid in the basin and the slush/warming equipment. This could pose a risk of infection to patients receiving warmed irrigation fluid or surgical slush if used during a procedure.
- Root cause
- Process control
- Action
- The firm sent a recall communication letter "Urgent: MEDICAL DEVICE RECALL IMMEDIATE ATTENTION REQUIRED" to customers on 01/11/2021.
- Quantity
- 132 cases (3168 drapes)
- Distribution
- Worldwide distribution - US Nationwide Distribution in the states of AL, AZ, CA, CO, FL, HI, ID, IL, IN, KS, LA, MA, MD, MI, MO, MT, NJ, NY, OH, OK, OR, PA, TN, TX, UT, WA. In the country of Japan.
- Lot, serial or model codes
- Model ORS-320, lot D192619A, 132 cases (3168 drapes)
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- PUI
- Firm FEI
- 3014342658
- Firm city
- Saint Paul
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28