Skip to the content

FDA Medical Device Recalls (openFDA)

Ecolab Inc: Slush + Warmer Disc Drape; Model ORS-320 Slush + Warmer Disc Drapes - Product Usage: single-use equipment cover intended for use during various surgeries where warm irrigation, slush and/or cold solution is required.

Recall number
Z-1141-2021
Recalling firm
Ecolab Inc
Product
Slush + Warmer Disc Drape; Model ORS-320 Slush + Warmer Disc Drapes - Product Usage: single-use equipment cover intended for use during various surgeries where warm irrigation, slush and/or cold solution is required.
Status
Terminated
Initiated
2021-01-11
Posted
not on file
Terminated
2023-01-09
Reason
Slits were identified in a low quantity of ORS-320 Slush + Warmer Disc Drapes from a single lot. A source for potential manufacturing error was identified and investigation is underway to prevent further recurrence. The slits may lead to a breach in the sterile barrier between fluid in the basin and the slush/warming equipment. This could pose a risk of infection to patients receiving warmed irrigation fluid or surgical slush if used during a procedure.
Root cause
Process control
Action
The firm sent a recall communication letter "Urgent: MEDICAL DEVICE RECALL IMMEDIATE ATTENTION REQUIRED" to customers on 01/11/2021.
Quantity
132 cases (3168 drapes)
Distribution
Worldwide distribution - US Nationwide Distribution in the states of AL, AZ, CA, CO, FL, HI, ID, IL, IN, KS, LA, MA, MD, MI, MO, MT, NJ, NY, OH, OK, OR, PA, TN, TX, UT, WA. In the country of Japan.
Lot, serial or model codes
Model ORS-320, lot D192619A, 132 cases (3168 drapes)
510(k) numbers
not on file
PMA numbers
not on file
Product code
PUI
Firm FEI
3014342658
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register