Siemens Healthcare Diagnostics, Inc.: Dimension Vista Ammonia Flex reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the quantitative measurement of ammonia in human plasma on the Dimension clinical chemistry system. Ammonia measurements are used in the diagnosis and treatment of severe liver disorders such as cirrhosis, hepatitis and Reye s syndrome
- Recall number
- Z-1141-2017
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Dimension Vista Ammonia Flex reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the quantitative measurement of ammonia in human plasma on the Dimension clinical chemistry system. Ammonia measurements are used in the diagnosis and treatment of severe liver disorders such as cirrhosis, hepatitis and Reye s syndrome
- Status
- Terminated
- Initiated
- 2016-12-16
- Posted
- not on file
- Terminated
- 2018-01-09
- Reason
- Siemens Healthcare Diagnostics has determined that Dimension Vista AMM (K3119) Flex reagent cartridge flex lots 16187BE, 16225BB, 16265AB do not meet the 60-day calibration interval claim due to reagent instability and results may show an Abnormal Assay.
- Root cause
- Under Investigation by firm
- Action
- Siemens sent an Urgent Field Safety Notice on December 20, 2016, to all customers who were affected by the issue. The letter identified the reason for the correction, the associated risks to health, and actions to be taken by the customer. Customers were instructed to discontinue use of and discard the affected lots, complete and return the Field Correction Effectiveness Check Form attached to the letter. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens technical support representative. For questions regarding this recall call 800441-9250.
- Quantity
- 8,032 units
- Distribution
- Worldwide Distribution - US (Nationwide) and Internationally to Argentina, Austria, Australia, Bangladesh, Belgium, Brazil, Bosnia Herzegovina. ,Canada, Canary Islands, Chile, China, Colombia, Croatia, Denmark, Ecuador, Egypt, France, Germany, Great Britain, Hong Kong, Hungary Israel ,India, Indonesia, Italy, Lebanon, Malaysia, Mexico, Nicaragua, Netherland, New Zealand, Philippines, Poland, Portugal, Republic of Korea, Saudi Arabia, Serbia, Slovakia, Slovenia, Spain, Sri Lanka, South Africa, Switzerland, Taiwan, Turkey, Uruguay, United.Arab Emirates, Venezuela, Yemen, and Japan.
- Lot, serial or model codes
- lots 16187BE, 16225BB and 16265AB
- 510(k) numbers
- ["K123320"]
- PMA numbers
- not on file
- Product code
- JIF
- Firm FEI
- 2517506
- Firm city
- Newark
- Firm state
- DE
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28