Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Healthcare Informatics, Inc.: iSite PACS image management system; Picture Archiving and Communications System; Radiology software package, Philips Healthcare Informatics, Foster City, CA 94404.

Recall number
Z-1141-2008
Recalling firm
Philips Healthcare Informatics, Inc.
Product
iSite PACS image management system; Picture Archiving and Communications System; Radiology software package, Philips Healthcare Informatics, Foster City, CA 94404.
Status
Terminated
Initiated
2008-01-04
Posted
2008-05-08
Terminated
2009-10-07
Reason
Incorrect display of one of the patient's images. This malfunction could cause an image to be incorrectly displayed, for example flipped left to right and anterior to posterior. The original image is still presented in the correct orientation next to the incorrect image when displayed on the monitor.
Root cause
Software design
Action
Philips Healthcare Informatics mailed customers notification: Urgent Product Correction regarding FCO#2954704-01/04/08-001-C. Firm released a corrected version of software (iSite PACS v.3.5.69) on January 21, 2008. Firm is scheduling customers to upgrade to the corrected version.
Quantity
197 units
Distribution
Worldwide Distribution: USA, United Arab Emirates, Germany, the Netherlands, Denmark, Australia, France, Italy, Belgium, Ireland, Italy, and Canada.
Lot, serial or model codes
iSITE PACS version 3.5x (Versions 3.5 and up to and including 3.5.65): 3.5.0.27, 3.5.0.36, 3.5.36.4, 3.5.38, 3.5.51.0, 3.5.51.02, 3.5.51.05, 3.5.56.0, 3.5.57.1, 3.5.57.2, 3.5.57.3, 3.5.57.4, and  3.5.65.0
510(k) numbers
["K042292"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3002971142
Firm city
Foster City
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register