Stryker Corporation: stryker Precision Thin, Catalog Number REF 2296-003-108; Micro Oscillating and Sagittal Blades

Recall number
Z-1138-2025
Recalling firm
Stryker Corporation
Product
stryker Precision Thin, Catalog Number REF 2296-003-108; Micro Oscillating and Sagittal Blades
Status
Open, Classified
Initiated
2025-01-08
Posted
2025-02-11
Terminated
not on file
Reason
Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
Root cause
Process control
Action
Stryker issued an URGENT MEDICAL DEVICE recall notice to its consignees on 01/08/2025 via FedEx. The notice explained the issue, potential risk, and requested the affected items be quarantined pending return. Stryker will issue a shipping label to the consignee once the response form has been received. Distributors were directed to forward the notice to all those to whom the affected devices were distributed.
Quantity
622 units
Distribution
Worldwide distribution.
Lot, serial or model codes
GTIN 04546540046482, Lot Number 22298017
510(k) numbers
["K143145"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3015967359
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register