Skip to the content

FDA Medical Device Recalls (openFDA)

Topcon Medical Systems, Inc.: IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.

Recall number
Z-1138-2019
Recalling firm
Topcon Medical Systems, Inc.
Product
IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.
Status
Terminated
Initiated
2019-03-06
Posted
2019-04-12
Terminated
2019-07-17
Reason
When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.
Root cause
Labeling design
Action
Topcon issued Device Correction Notice on 3/22/19 providing problem- there exists a situation in which the displayed patient name and the thumbnail image do not match and action to take.Topcon has corrected this issue in IMAGEnet 6 version 2.0.3 and will be providing a remote upgrade free of charge. Contact customer support team to schedule the upgrade at 866.922.6278 option 1 or by email at TMSservice@topcon.com
Quantity
450 units
Distribution
Nationwide
Lot, serial or model codes
System Version 1.53
510(k) numbers
["K171370"]
PMA numbers
not on file
Product code
NFJ
Firm FEI
2242863
Firm city
Oakland
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register