FDA Medical Device Recalls (openFDA)
Endogastric Solutions Inc: EndoGastric Solutions EsophyX2 Plus Device with Serosa Fastener and Accessories Part number C02033-01. Label reads in part "EsophyX2 2-LINK Device Qty: (1) Fastener Delivery System ***" Designed for the treatment of Gastroesophageal Reflux Disease (GERD), with erosive and non-erosive esophagitis
- Recall number
- Z-1136-2013
- Recalling firm
- Endogastric Solutions Inc
- Product
- EndoGastric Solutions EsophyX2 Plus Device with Serosa Fastener and Accessories Part number C02033-01. Label reads in part "EsophyX2 2-LINK Device Qty: (1) Fastener Delivery System ***" Designed for the treatment of Gastroesophageal Reflux Disease (GERD), with erosive and non-erosive esophagitis
- Status
- Terminated
- Initiated
- 2013-02-04
- Posted
- 2013-04-17
- Terminated
- 2013-07-02
- Reason
- EsophyX2Plus device was not validated under actual or simulated use conditions prior to distributing the product to physicians.
- Root cause
- Release of Material/Component prior to receiving test results
- Action
- EndoGastric Solutions, Inc. sent an Urgent Medical Device Recall letter dated January 31, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to segregate and record lot numbers for any affected product. If no product remain in inventory customers were instructed to report such to customer service. Upon receipt of the product a credit will be issued for each device returned. For any questions customers were instructed to call 425-307-9226. For questions regarding this recall call 425-307-9200.
- Quantity
- 190
- Distribution
- Nationwide Distribution including NC, VA, WI, TN, KY, IN, FL, MI, IA, IL, TX, LA, AR, HI, CA, NV, OR, and WA,
- Lot, serial or model codes
- Lots 401378, 401392, 401409, and 401416 Additional lots manufactured from 11/26/12 to 12/07/12 had a stop ship issued issued on 11/21/12 and the product was placed in quarantine (it was not distributed). This represents an additional 175 units from a total of 187 units manufactured during this time period. 12 units were used for bench testing. The lot numbers of units in quarantine are 401432, 401434, 401436, and 401441.
- 510(k) numbers
- ["K092400"]
- PMA numbers
- not on file
- Product code
- ODE
- Firm FEI
- 3005473391
- Firm city
- Redmond
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28