Skip to the content

FDA Medical Device Recalls (openFDA)

Endogastric Solutions Inc: EndoGastric Solutions EsophyX2 Plus Device with Serosa Fastener and Accessories Part number C02033-01. Label reads in part "EsophyX2 2-LINK Device Qty: (1) Fastener Delivery System ***" Designed for the treatment of Gastroesophageal Reflux Disease (GERD), with erosive and non-erosive esophagitis

Recall number
Z-1136-2013
Recalling firm
Endogastric Solutions Inc
Product
EndoGastric Solutions EsophyX2 Plus Device with Serosa Fastener and Accessories Part number C02033-01. Label reads in part "EsophyX2 2-LINK Device Qty: (1) Fastener Delivery System ***" Designed for the treatment of Gastroesophageal Reflux Disease (GERD), with erosive and non-erosive esophagitis
Status
Terminated
Initiated
2013-02-04
Posted
2013-04-17
Terminated
2013-07-02
Reason
EsophyX2Plus device was not validated under actual or simulated use conditions prior to distributing the product to physicians.
Root cause
Release of Material/Component prior to receiving test results
Action
EndoGastric Solutions, Inc. sent an Urgent Medical Device Recall letter dated January 31, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to segregate and record lot numbers for any affected product. If no product remain in inventory customers were instructed to report such to customer service. Upon receipt of the product a credit will be issued for each device returned. For any questions customers were instructed to call 425-307-9226. For questions regarding this recall call 425-307-9200.
Quantity
190
Distribution
Nationwide Distribution including NC, VA, WI, TN, KY, IN, FL, MI, IA, IL, TX, LA, AR, HI, CA, NV, OR, and WA,
Lot, serial or model codes
Lots 401378, 401392, 401409, and 401416 Additional lots manufactured from 11/26/12 to 12/07/12 had a stop ship issued issued on 11/21/12 and the product was placed in quarantine (it was not distributed). This represents an additional 175 units from a total of 187 units manufactured during this time period. 12 units were used for bench testing. The lot numbers of units in quarantine are 401432, 401434, 401436, and 401441.
510(k) numbers
["K092400"]
PMA numbers
not on file
Product code
ODE
Firm FEI
3005473391
Firm city
Redmond
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register