FDA Medical Device Recalls (openFDA)
Edwards Lifesciences, LLC: Edwards Lifesciences Crimpers, Model 9100CR26. The Crimpers are used to crimp the transcatheter heart valve onto the delivery system prior to use.
- Recall number
- Z-1135-2013
- Recalling firm
- Edwards Lifesciences, LLC
- Product
- Edwards Lifesciences Crimpers, Model 9100CR26. The Crimpers are used to crimp the transcatheter heart valve onto the delivery system prior to use.
- Status
- Terminated
- Initiated
- 2013-03-26
- Posted
- 2013-04-17
- Terminated
- 2013-09-16
- Reason
- Holes in the sterile barrier pouch may be caused by handling of the device during manufacturing or once pouches have been removed from the unit box and stored on the shelf by customers.
- Root cause
- Package design/selection
- Action
- Edwards Lifesciences notified customers via letter and instructed them to maintain the device in its protective unit carton until ready for use and to inspect the pouch for holes or tears prior to use as instructed in the product's labeling. If any holes or tears are found, customers are instructed to contact Edwards Customer Service to arrange for return and replacement of damaged product.
- Quantity
- 6,117 units
- Distribution
- Nationwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI.
- Lot, serial or model codes
- Models 9100CR26, all lot numbers.
- 510(k) numbers
- not on file
- PMA numbers
- ["P100041","P110021"]
- Product code
- NPT
- Firm FEI
- 1713910
- Firm city
- Draper
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28