Skip to the content

FDA Medical Device Recalls (openFDA)

Edwards Lifesciences, LLC: Edwards Lifesciences Crimpers, Model 9100CR26. The Crimpers are used to crimp the transcatheter heart valve onto the delivery system prior to use.

Recall number
Z-1135-2013
Recalling firm
Edwards Lifesciences, LLC
Product
Edwards Lifesciences Crimpers, Model 9100CR26. The Crimpers are used to crimp the transcatheter heart valve onto the delivery system prior to use.
Status
Terminated
Initiated
2013-03-26
Posted
2013-04-17
Terminated
2013-09-16
Reason
Holes in the sterile barrier pouch may be caused by handling of the device during manufacturing or once pouches have been removed from the unit box and stored on the shelf by customers.
Root cause
Package design/selection
Action
Edwards Lifesciences notified customers via letter and instructed them to maintain the device in its protective unit carton until ready for use and to inspect the pouch for holes or tears prior to use as instructed in the product's labeling. If any holes or tears are found, customers are instructed to contact Edwards Customer Service to arrange for return and replacement of damaged product.
Quantity
6,117 units
Distribution
Nationwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI.
Lot, serial or model codes
Models 9100CR26, all lot numbers.
510(k) numbers
not on file
PMA numbers
["P100041","P110021"]
Product code
NPT
Firm FEI
1713910
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register