FDA Medical Device Recalls (openFDA)
GE Healthcare: GE Healthcare Definium 5000 Digital Radiographic Imaging System. Model numbers 5220493. The Definium 5000 System is designed to handle radiographic applications using the digital system. The system configuration includes a choice of elevating or non-elevating or no table, a floor column stand with rotating U-arm, and GE's patented Digital detector that captures radiographic images in digital form, as well as an X-ray generator/power unit.
- Recall number
- Z-1133-2008
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Definium 5000 Digital Radiographic Imaging System. Model numbers 5220493. The Definium 5000 System is designed to handle radiographic applications using the digital system. The system configuration includes a choice of elevating or non-elevating or no table, a floor column stand with rotating U-arm, and GE's patented Digital detector that captures radiographic images in digital form, as well as an X-ray generator/power unit.
- Status
- Terminated
- Initiated
- 2007-12-17
- Posted
- 2008-09-04
- Terminated
- 2008-09-10
- Reason
- GE Definium 5000 Collimator failed compliance testing due to a blade sizing issue.
- Root cause
- Radiation Control for Health and Safety Act
- Action
- A consignee letter dated 12/12/07 was sent to consignees (Director or Manager of Radiological Imaging, Clinical Administrator). The letter stated that an installed collimator failed compliance testing because of a blaze sizing issue. A GE Field Engineer site visit was also conducted (FMI No. 10814 12/18/07) to implement the correction. Contact GE Healthcare at 1-262-544-3894.
- Quantity
- 14 devices (12US & 2 OUS)
- Distribution
- MD, FL, AZ, TX, MS, NY, OH, CA, GERMANY, and FRANCE.
- Lot, serial or model codes
- Model Number 5220493: PC0127XR06, M1973721
- 510(k) numbers
- ["K063283"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28