FDA Medical Device Recalls (openFDA)
Medtronic Perfusion Systems: Cardioblate" Gemini"-s Irrigated RF Surgical Ablation System; The device is placed in a plastic tray with a snap on plastic lid, the tray is placed in protective wrap, then into a sterile pouch, 1 device per carton, 1 carton per box.
- Recall number
- Z-1131-2022
- Recalling firm
- Medtronic Perfusion Systems
- Product
- Cardioblate" Gemini"-s Irrigated RF Surgical Ablation System; The device is placed in a plastic tray with a snap on plastic lid, the tray is placed in protective wrap, then into a sterile pouch, 1 device per carton, 1 carton per box.
- Status
- Open, Classified
- Initiated
- 2022-04-20
- Posted
- not on file
- Terminated
- not on file
- Reason
- Firm detected an increase in complaints related to fractured jaw tips of the Cardioblate Gemini-s.
- Root cause
- Labeling design
- Action
- A recall notification was sent to consignees on 04/20/2022 via the United Parcel Service. The letter instructs the consignees to review the updated guidance on device usage contained in the letter, share the notice to those that need to be aware, and complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com.
- Quantity
- 1236 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, ND, MA, ID, WI, KS, UT, MN, NY, NJ, GA, MO and the countries of Australia, Belgium, China, Czech Republic, France, Germany, Greece, Israel, Italy, Kazakhstan, Latvia, Netherlands, Poland, Russian Federation, Singapore, Switzerland.
- Lot, serial or model codes
- ABLATION DEVICE 49260 PROCEDURE KIT Model 49260; GTIN: 00613994268051, Lot Numbers: 215G, 256E, 388C, 466C, 584D, 611A, 745F, 840C, 850C, 967C; GTIN: 00613994755643, Lot Numbers: 469C, 612A, 972C, 973C, GTIN: 00643169983120, Lot Numbers: 125E, 238D, 257E, 258E, 259E, 335D, 373B, 410E, 428D, 454F, 467C, 585D, 632E, 635F, 700D, 721C, 722C, 732E, 735E, 738E, 841C, 876E, 923D, 968C, 972E, 980E, 995D. CLAMP 49351 CARDIOBLATE GEMINI-S FT Model: 49351; GTIN: 00643169998117, Lot Numbers: 495F, 734E, 846D.
- 510(k) numbers
- ["K121767","K200514"]
- PMA numbers
- not on file
- Product code
- OCL
- Firm FEI
- 1000116158
- Firm city
- Brooklyn Park
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28