Biomet, Inc.: Central Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder prosthesis assembly and/or fixation.
- Recall number
- Z-1131-2021
- Recalling firm
- Biomet, Inc.
- Product
- Central Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder prosthesis assembly and/or fixation.
- Status
- Terminated
- Initiated
- 2021-01-15
- Posted
- not on file
- Terminated
- 2023-03-09
- Reason
- Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.
- Root cause
- Labeling Change Control
- Action
- On 01/19/2019, Zimmer Biomet issued Urgent Medical Device Recall notices to distributors with direct sales and/or fielded inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory will be notified via FedEx. Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.
- Quantity
- 49
- Distribution
- US Nationwide distribution including in the states of MS, MN, KY, FL, MI, OH, AZ, IL, WI, CA, UT, TN.
- Lot, serial or model codes
- Model No.: 115395, Lot No.: 077170, UDI: (01) 00880304677074 (17) 301010 (10) 077170
- 510(k) numbers
- ["K132239"]
- PMA numbers
- not on file
- Product code
- PHX
- Firm FEI
- 1825034
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28