Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Manufacturing B.V.: Product 33 consists of all product under product code: JDI and same usage: Item no: 662406515 HGP II ACETABULAR CUP BON 662406520 HGP II ACETABULAR CUP BON 662406525 HGP II ACETABULAR CUP BON 662406530 HGP II ACETABULAR CUP BON 662406535 HGP II ACETABULAR CUP BON 662406540 HGP II ACETABULAR CUP BON 662406550 HGP II ACETABULAR CUP BON 662406560 HGP II ACETABULAR CUP BON For use in total hip arthroplasty

Recall number
Z-1131-2016
Recalling firm
Zimmer Manufacturing B.V.
Product
Product 33 consists of all product under product code: JDI and same usage: Item no: 662406515 HGP II ACETABULAR CUP BON 662406520 HGP II ACETABULAR CUP BON 662406525 HGP II ACETABULAR CUP BON 662406530 HGP II ACETABULAR CUP BON 662406535 HGP II ACETABULAR CUP BON 662406540 HGP II ACETABULAR CUP BON 662406550 HGP II ACETABULAR CUP BON 662406560 HGP II ACETABULAR CUP BON For use in total hip arthroplasty
Status
Terminated
Initiated
2016-01-11
Posted
2016-03-12
Terminated
2017-04-13
Reason
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root cause
Packaging
Action
Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
Quantity
2236
Distribution
US Nationwide Distribution
Lot, serial or model codes
lot no.: 61674472 61906696 62016799 62332483 61674473 61902194 61906697 62326386 61674476 61953546 62161131 62172656 62326385 61953547 62011539 62017894 62130757 61886072 61953548 62017895 62172659 61886073 61948763 61953551 62172661 61874793 61948764 61996232 62017897 62172663 61874794 61953554 62286122
510(k) numbers
["K840643"]
PMA numbers
not on file
Product code
JDI
Firm FEI
1000132037
Firm city
Mercedita
Firm state
not on file
Firm country
PR

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register