Skip to the content

FDA Medical Device Recalls (openFDA)

HeartWare Inc: HeartWare Ventricular Assist System (HeartWare Controller) Product Usage: For use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare System is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter.

Recall number
Z-1131-2015
Recalling firm
HeartWare Inc
Product
HeartWare Ventricular Assist System (HeartWare Controller) Product Usage: For use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare System is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter.
Status
Terminated
Initiated
2015-01-16
Posted
2015-02-25
Terminated
2020-02-05
Reason
The affected clinical trial Controllers exhibit a higher susceptibility to ESD than newer commercial Controllers.
Root cause
Device Design
Action
HeartWare sent an URGENT MEDICAL DEVICE RECALL dated January 16, 2015 asking clinicians to quarantine the affected controllers and complete and return the acknowledgement form. HeartWare also asks clinicians to identify patients with affected controllers and review the applicable risks with the patient and as soon as reasonably possible if medically advisable, exchange the affected controller under medical supervision with a new controller. HeartWare also asks that for each patient using an affected controller as their backup to contact the patient and arrange to have the backup controller replaced with a new controller. Lastly, HeartWare asks the firm to complete the patient tracking form and return all quarantined controllers to HeartWare along with the completion form. Should you have any questions or concerns, please contact your local HeartWare representative. Our 24-Hour Clinical Support personnel are also available at 1-888-494-6365.
Quantity
4,845
Distribution
Worldwide Distribution - US Nationwide in the states of: AL, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NY, OH, OR, PA, TX, UT, VA, WA and District of Columbia and countries of: Canada, Poland, Australia, Austria, Italy, DE, New Zealand, South Africa, France, Switzerland, Germany, Brazil, UK, Belgium, Netherlands, Israel, Malaysia, Turkey, Greece, Japan, Singapore, Poland, Lithuania, Sweden, Norway, Finland, and India.
Lot, serial or model codes
Product Codes- US product: Model No.: 1400, 1401US; International product: Model No.: 1400, 1401XX ('XX' represents country designation) Serial No.: CON000001 through CON005472
510(k) numbers
not on file
PMA numbers
["P100047"]
Product code
DSQ
Firm FEI
3007042319
Firm city
Miami Lakes
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register