Possis Medical, Inc: Possis AngioJet XVG Thrombectomy Sets, Model/Order # 105042-001, Sterilized with Ethylene Oxide, Possis Medical, Inc., 9055 Evergreen Boulevard NW, Minneapolis, MN 55433-8003. (These sets are used with the new AngioJet Ultra Console)

Recall number
Z-1131-2007
Recalling firm
Possis Medical, Inc
Product
Possis AngioJet XVG Thrombectomy Sets, Model/Order # 105042-001, Sterilized with Ethylene Oxide, Possis Medical, Inc., 9055 Evergreen Boulevard NW, Minneapolis, MN 55433-8003. (These sets are used with the new AngioJet Ultra Console)
Status
Terminated
Initiated
2007-06-19
Posted
2007-08-10
Terminated
2008-03-02
Reason
Mislabeling : Possis AngioJet XVG Catheter Thrombectomy Sets display incorrect information on the outer packaging box and also on the inner sterile tray. The labeling incorrectly states that the unit is indicated for use in coronary vessels. These units should indicated use only for peripheral arteries.
Root cause
Other
Action
A Urgent Medical Device recall letter dated 19 June 2007, was sent to consignees advising that all unused product be removed and returned to Possis. An acknowledgement of the disposiiton of the product will be requested from consignees.
Quantity
194
Distribution
AR, AZ, CA, FL, GA, IL, IN, KS, KY, MA, ME, MI, MN, MO, MS, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, WA, WI, WV
Lot, serial or model codes
71851, 71852, 71997, 72086, 72539, 73170, 73171, 73646, 73758.
510(k) numbers
["K050794"]
PMA numbers
not on file
Product code
DXE
Firm FEI
2183460
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register