Skip to the content

FDA Medical Device Recalls (openFDA)

Cardinal Health 200, LLC: Cardinal Health Presource Kits: 1) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF104; 2) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF105 3) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF106 4) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF107 5) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF108 6) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF109 7) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10V 8) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10W 9) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10Y 10) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10Z 11) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF111 12) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF112 13) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCP2 14) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCP3 15) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCP4 16) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPA 17) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPB 18) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPC 19) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPD 20) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCPE 21) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCPF

Recall number
Z-1130-2025
Recalling firm
Cardinal Health 200, LLC
Product
Cardinal Health Presource Kits: 1) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF104; 2) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF105 3) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF106 4) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF107 5) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF108 6) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF109 7) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10V 8) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10W 9) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10Y 10) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10Z 11) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF111 12) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF112 13) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCP2 14) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCP3 15) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCP4 16) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPA 17) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPB 18) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPC 19) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPD 20) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCPE 21) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCPF
Status
Open, Classified
Initiated
2024-11-08
Posted
2025-02-10
Terminated
not on file
Reason
Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.
Root cause
Material/Component Contamination
Action
An URGENT: MEDICAL DEVICE CORRECTION OVER LABEL notification letter dated 11/8/24 was sent to customers. If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all CODMAN Patties & Strips included with the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.
Quantity
973,785 total units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia.
Lot, serial or model codes
1) PWV2MF104, UDI-DI:10888439676167 (each) 50888439676165 (case), Lot Number: J46C10 ; 2) PWV2MF105, UDI-DI:10195594004217 (each) 50195594004215 (case), Lot Number: J8998R ; 3) PWV2MF106, UDI-DI:10195594095994 (each) 50195594095992 (case), Lot Number: 5136000 ; 4) PWV2MF107, UDI-DI:10195594148065 (each) 50195594148063 (case), Lot Number: 5140389 ; 5) PWV2MF108, UDI-DI:10195594228941 (each) 50195594228949 (case), Lot Number: 5140389 ; 6) PWV2MF109, UDI-DI:10195594394875 (each) 50195594394873 (case), Lot Number: J65K18 ; 7) PWV2MF10V, UDI-DI:10888439342741 (each) 50888439342749 (case), Lot Number: HP1027 ; 8) PWV2MF10V, UDI-DI:10888439342741 (each) 50888439342749 (case), Lot Number: J4220A ; 9) PWV2MF10V, UDI-DI:10888439342741 (each) 50888439342749 (case), Lot Number: J43W15 ; 10) PWV2MF10V, UDI-DI:10888439342741 (each) 50888439342749 (case), Lot Number: J4953C ; 11) PWV2MF10V, UDI-DI:10888439342741 (each) 50888439342749 (case), Lot Number: J6022K ; 12) PWV2MF10V, UDI-DI:10888439342741 (each) 50888439342749 (case), Lot Number: J6372Z ; 13) PWV2MF10V, UDI-DI:10888439342741 (each) 50888439342749 (case), Lot Number: J7708T ; 14) PWV2MF10W, UDI-DI:10888439693645 (each) 50888439693643 (case), Lot Number: J8241D ; 15) PWV2MF10Y, UDI-DI:10888439967432 (each) 50888439967430 (case), Lot Number: J84K11 ; 16) PWV2MF10Y, UDI-DI:10888439967432 (each) 50888439967430 (case), Lot Number: J84K10 ; 17) PWV2MF10Z, UDI-DI:10195594180294 (each) 50195594180292 (case), Lot Number: J65K19 ; 18) PWV2MF10Z, UDI-DI:10195594180294 (each) 50195594180292 (case), Lot Number: 5775812 ; 19) PWV2MF10Z, UDI-DI:10195594180294 (each) 50195594180292 (case), Lot Number: 6116189 ; 20) PWV2MF10Z, UDI-DI:10195594180294 (each) 50195594180292 (case), Lot Number: 6251076 ; 21) PWV2MF111, UDI-DI:10195594634957 (each) 50195594634955 (case), Lot Number: 5858469 ; 22) PWV2MF112, UDI-DI:10195594711177 (each) 50195594711175 (case), Lot Number: 6022146 ; 23) PWV2MFCP2, UDI-DI:10197106368718 (each) 50197106368716 (case), Lot Number: 6788690 ; 24) PWV2MFCP3, UDI-DI:10197106738399 (each) 50197106738397 (case), Lot Number: 7333448 ; 25) PWV2MFCP4, UDI-DI:10197106784167 (each) 50197106784165 (case), Lot Number: 7360541 ; 26) PWV2MFCPA, UDI-DI:10197106143469 (each) 50197106143467 (case), Lot Number: 6701855 ; 27) PWV2MFCPB, UDI-DI:10197106245491 (each) 50197106245499 (case), Lot Number: 6702040 ; 28) PWV2MFCPB, UDI-DI:10197106245491 (each) 50197106245499 (case), Lot Number: 6788692 ; 29) PWV2MFCPC, UDI-DI:10197106425701 (each) 50197106425709 (case), Lot Number: 6788692 ; 30) PWV2MFCPD, UDI-DI:10197106484340 (each) 50197106484348 (case), Lot Number: 7160517 ; 31) PWV2MFCPD, UDI-DI:10197106484340 (each) 50197106484348 (case), Lot Number: 7160540 ; 32) PWV2MFCPE, UDI-DI:10197106532461 (each) 50197106532469 (case), Lot Number: 7287574 ; 33) PWV2MFCPF, UDI-DI:10197106548059 (each) 50197106548057 (case), Lot Number: 7287574 ; 34) PWV2MFCPF, UDI-DI:10197106548059 (each) 50197106548057 (case), Lot Number: 7287577 ; 35) PWV2MFCPF, UDI-DI:10197106548059 (each) 50197106548057 (case), Lot Number: 7331480 ;
510(k) numbers
not on file
PMA numbers
not on file
Product code
OFX
Firm FEI
3001236905
Firm city
Waukegan
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register