FDA Medical Device Recalls (openFDA)

Haemonetics Corporation: Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.

Recall number
Z-1130-2015
Recalling firm
Haemonetics Corporation
Product
Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.
Status
Terminated
Initiated
2015-01-26
Posted
2015-02-19
Terminated
2015-06-05
Reason
Specific lots of Level II Control Kits produce high Maximum Amplitude (MA) results, which are outside of the manufacturer's expected Quality Control range for MA parameter.
Root cause
Process control
Action
Haemonetics issued on 1/26/15, an Urgent Medical Device Recall to consignees The letter identified the affected product, the reason for the recall, as well as the risk to health. It instructs consignees to inspect their stock and determine if they have any of the affected lots. Customers are to immediately discontinue use and discard as instructed any affected product. The attached customer acknowledgment form should be completed and returned whether or not customers have any affected product. Customers are to contact their local service representative at 1-800-537-2802 if replacement product is needed. If any affected product has been forwarded to other locations, customers are to inform them of the recall. Please call the Technical Support team at 1-800-438-2834, with questions about this action.
Quantity
4,143 kits
Distribution
Worldwide Distribution -- US, Canada, Czech Republic, Denmark, Great Britain, and Sweden.
Lot, serial or model codes
Lot numbers: HMO 9200  HMO 9217  HMO 9226  HMO 9232  HMO 9276  HMO 9285  HMO 9292 Exp. Date 1/2017
510(k) numbers
["K002177"]
PMA numbers
not on file
Product code
JPA
Firm FEI
1219343
Firm city
Braintree
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register