Skip to the content

FDA Medical Device Recalls (openFDA)

Djo Surgical: FMP X-alt Acetabular Liner This acetabular liner is intended for use with DJO Surgical hip systems. It is used with an acetabular shell to replicate the kinematics of a natural hip.

Recall number
Z-1130-2014
Recalling firm
Djo Surgical
Product
FMP X-alt Acetabular Liner This acetabular liner is intended for use with DJO Surgical hip systems. It is used with an acetabular shell to replicate the kinematics of a natural hip.
Status
Terminated
Initiated
2014-02-07
Posted
2014-03-04
Terminated
2014-10-01
Reason
Packaging error -- two different types and sizes of acetabular liners used during hip replacement surgery were mistakenly packaged into each other's outer boxes after sterilization, resulting in the products being mislabeled. The units labeled as Model/Catalog Number 931-28-248 Lot/Serial Number 685F1034 on the outside and containing Model/Catalog Number 932-36-252, Lot# 728F1089 on the inside (1
Root cause
Packaging process control
Action
DJO Global sent an Urgent Field Safety Notice dated February 7, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. All 10 devices with the outer label of 931-28-248, Lot#685F1034 were found in-house and were quarantined. The agencies who have the affected product (devices labeled as 932-36-252, Lot# 728F1089 containing 931-28-248, Lot#685F1034) were contacted to quarantine their devices. Of the affected devices, 2 of the lot of 10 have already been returned via product complaint. Customers were instructed to pass on the notice to all those who need to be aware within their organization or to any organization where the potentially affected devices have been transferred. Customers were instructed to place a replacement order and receive an RMA number to return all affected devices.. Customers with questions were instructed to call 512-834-6255.
Quantity
10 units of Model/Catalog Number 931-28-248 Lot/Serial Number(s) 685F1034 labeled asModel/Catalog Number 932-36-252, Lot# 728F1089
Distribution
Nationwide Distribution - USA including TX, LA, FL, CA, and AZ.
Lot, serial or model codes
Model/Catalog Number 931-28-248 Lot/Serial Number 685F1034 (an MP5 Non-Hooded, Neutral Liner)  were mistakenly packaged/labeled as  Model/Catalog Number 932-36-252, Lot# 728F1089 (an MP7 10-degree Hooded Liner)  and vice-versa
510(k) numbers
["K072154"]
PMA numbers
not on file
Product code
LPH
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register