FDA Medical Device Recalls (openFDA)
Alere San Diego: Product Brand Names: Triage Drugs of Abuse Panel 8 Test Kit (K924476), Triage Drugs of Abuse Panel Plus TCA (K973784), & Triage Drugs of Abuse Panel Plus PPX (K014247) Model Number: 90008, 92002, & 92000 Type of Packaging (i.e. box, plastic bag): For all affected device types: 25 individually pouched devices in a labeled kit box. The Triage Drugs of Abuse Panel plus Tricyclic Antidepressants utilizes a patented immunochemistry, ASCENDTM MULTIMMUNOASSAY (AMIATM). Each AMIATM assay is a competitive binding immunoassay in which a chemically labeled drug (drug conjugate) competes with drug which may be present in the urine for antibody binding sites. After a brief incubation, the reaction mixture is transferred to the membrane in the Detection Area. Free drug conjugate that is displaced from antibody binding sites by drug in the urine, binds to a zone of monoclonal antibody that is immobilized on the membrane. The membrane is washed to remove the unbound conjugate and clear the background. Test results are visually read. The Triage Drugs of Abuse Panel plus Tricyclic Antidepressants is an immunoassay used for the qualitative determination of the presence of the major metabolites of drugs of abuse, (Phencyclidine, Benzodiazepines, Cocaine Metabolite, Amphetamines, THC, Opiates, Barbiturates), and Tricyclic Antidepressants in urine.
- Recall number
- Z-1130-2012
- Recalling firm
- Alere San Diego
- Product
- Product Brand Names: Triage Drugs of Abuse Panel 8 Test Kit (K924476), Triage Drugs of Abuse Panel Plus TCA (K973784), & Triage Drugs of Abuse Panel Plus PPX (K014247) Model Number: 90008, 92002, & 92000 Type of Packaging (i.e. box, plastic bag): For all affected device types: 25 individually pouched devices in a labeled kit box. The Triage Drugs of Abuse Panel plus Tricyclic Antidepressants utilizes a patented immunochemistry, ASCENDTM MULTIMMUNOASSAY (AMIATM). Each AMIATM assay is a competitive binding immunoassay in which a chemically labeled drug (drug conjugate) competes with drug which may be present in the urine for antibody binding sites. After a brief incubation, the reaction mixture is transferred to the membrane in the Detection Area. Free drug conjugate that is displaced from antibody binding sites by drug in the urine, binds to a zone of monoclonal antibody that is immobilized on the membrane. The membrane is washed to remove the unbound conjugate and clear the background. Test results are visually read. The Triage Drugs of Abuse Panel plus Tricyclic Antidepressants is an immunoassay used for the qualitative determination of the presence of the major metabolites of drugs of abuse, (Phencyclidine, Benzodiazepines, Cocaine Metabolite, Amphetamines, THC, Opiates, Barbiturates), and Tricyclic Antidepressants in urine.
- Status
- Terminated
- Initiated
- 2012-01-04
- Posted
- 2012-03-01
- Terminated
- 2013-07-05
- Reason
- Alere San Diego received a rise in customer complaints regarding false positive Barbiturates (BAR) results on the Triage Drugs of Abuse Panel Plus TCA.
- Root cause
- Nonconforming Material/Component
- Action
- Alere sent a "PRODUCT NOTIFICATION" letter dated January 4, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Verification Form was enclosed for customers to complete and return to the firm. Contact Technical Services at 1-877-308-8287.
- Quantity
- 693054 kits
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Australia, Brazil, Canada, EU, Guatemala, Japan, Lebanon, and Saudi Arabia.
- Lot, serial or model codes
- Part 90008: lot 259403 exp 02/03/13, lot 267081 exp 05/08/13 Part 92000: lots 254662, 255830, 257527, 262143 with exp 01/06/13; lots 259628, 259954 with exp 02/03/13 Part 92002: lot 234295 exp 02/03/13
- 510(k) numbers
- ["K924476","K973784"]
- PMA numbers
- not on file
- Product code
- DIS
- Firm FEI
- 1000125596
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28