Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: System, X-Ray, Tomography, Computed: Revolution Apex, Revolution CT with Apex Edition

Recall number
Z-1129-2021
Recalling firm
GE Healthcare, LLC
Product
System, X-Ray, Tomography, Computed: Revolution Apex, Revolution CT with Apex Edition
Status
Terminated
Initiated
2020-12-17
Posted
not on file
Terminated
2023-01-09
Reason
GE Healthcare has become aware of a potential issue on Revolution Apex, Revolution CT with Apex Edition, Revolution CT, and Revolution CT ES systems where the table pinch protector could be damaged or missing leading to exposed table screws which could cause a laceration due to a sharp edge.
Root cause
Process change control
Action
The manufacturer sent an "Urgent Medical Device Correction" letter (GEHC Ref# 25496) to customers with affected systems on 12/17/2020.
Quantity
96 units
Distribution
Worldwide - US Nationwide Distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NJ, NV, NY, NH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV; In the countries of Algeria, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Denmark, Estonia, Finland, France, Gabon, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea (Republic Of), Kuwait, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uzbekistan, Viet Nam.
Lot, serial or model codes
Model Numbers:  5590000-20: REV2A1900002CN, REV2A2000013CN, REV2A2000025CN, REV2A2000006CN, REV2A2000003CN, REV2A2000015CN, REV2A1900003CN, REV2A2000007CN, REV2A2000022CN, REV2A2000010CN, REV2A2000030CN, REV2A2000029CN, REV2A2000017CN, REV2A2000001CN, REV2A2000002CN, REV2A2000031CN, REV2A2000020CN, REV2A2000024CN, REV2A2000023CN, REV2A2000012CN, REV2A2000016CN, REV2A1900001CN, REV2A2000027CN, REV2A2000028CN, REV2A2000018CN, REV2A1900005CN, REV2A2000009CN, REV2A2000026CN, REV2A2000021CN, REV2A2000011CN  5590000-21: xxx, xxx, xxx* (66 units) *serial number information current as of 01/21/2021.
510(k) numbers
["K163213","K191777"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register