FDA Medical Device Recalls (openFDA)
Zimmer Manufacturing B.V.: Product 31 consists of all product under product code: KTT and same usage: Item no: 118113003 FREE-LOCK TUBE & SCP PLAT 118113502 FREE-LOCK TUBE & SCP PLAT 118113503 FREE-LOCK TUBE & SCP PLAT 118114002 FREE-LOCK TUBE & SCP PLAT 118114003 FREE-LOCK TUBE & SCP PLAT 118114502 FREE-LOCK TUBE & SCP PLAT 118114503 FREE-LOCK TUBE & SCP PLAT 118115002 FREE-LOCK TUBE & SCP PLAT 118115003 FREE-LOCK TUBE & SCP PLAT Product Usage: Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
- Recall number
- Z-1129-2016
- Recalling firm
- Zimmer Manufacturing B.V.
- Product
- Product 31 consists of all product under product code: KTT and same usage: Item no: 118113003 FREE-LOCK TUBE & SCP PLAT 118113502 FREE-LOCK TUBE & SCP PLAT 118113503 FREE-LOCK TUBE & SCP PLAT 118114002 FREE-LOCK TUBE & SCP PLAT 118114003 FREE-LOCK TUBE & SCP PLAT 118114502 FREE-LOCK TUBE & SCP PLAT 118114503 FREE-LOCK TUBE & SCP PLAT 118115002 FREE-LOCK TUBE & SCP PLAT 118115003 FREE-LOCK TUBE & SCP PLAT Product Usage: Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
- Status
- Terminated
- Initiated
- 2016-01-11
- Posted
- 2016-03-12
- Terminated
- 2017-04-13
- Reason
- As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
- Root cause
- Packaging
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
- Quantity
- 285
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- lot no.: 61771007 61827364 61853067 61865449 61870424 61888695 61953208 62230187 61853068 61870425 61953213 62230188 61860667 61895085 61937602 61953221 62230190 62271893 61900049 61900056 61958816 62124265 62124266 62241683 62255907 61778898 61860656 61870426 61853070 61900064 370958 61777470 61834885 61934150 370957 61888707 62271892
- 510(k) numbers
- ["K150818"]
- PMA numbers
- not on file
- Product code
- KTT
- Firm FEI
- 1000132037
- Firm city
- Mercedita
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28