Skip to the content

FDA Medical Device Recalls (openFDA)

Mako Surgical Corporation: Restoris Partial Knee Application (PKA) RIO (TGS 2.0). For use with the Robotic Arm Interactive Orthopedic System (RIO), is indicated for use in surgical knee procedures.

Recall number
Z-1128-2015
Recalling firm
Mako Surgical Corporation
Product
Restoris Partial Knee Application (PKA) RIO (TGS 2.0). For use with the Robotic Arm Interactive Orthopedic System (RIO), is indicated for use in surgical knee procedures.
Status
Terminated
Initiated
2015-01-22
Posted
2015-02-19
Terminated
2016-12-14
Reason
When using the MAKOplasty partial knee Arthroplasty application, the burr continues spinning outside of the stereotactic boundary and after the control switches (foot pedal and trigger) cease to be activated
Root cause
Device Design
Action
On January 22, 2015 the firm sent an IMPORTANT MEDICAL DEVICE CORRECTION asking consignees to complete the attached acknowledgement form and return them by email, fax or mail. The notice also noted the RIO system can continue to be used. The likelihood of occurrence is rare (.035% of cases) and the event is readily detectable and can be resolved. Should you have any questions regarding this Important Medical Device Recall Notice, to contact us at (954) 628-1721.
Quantity
228 units
Distribution
Nationwide Distribution-including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii.
Lot, serial or model codes
All lots, Part Numbers: 201000, 203999, 207300, and 209930.
510(k) numbers
["K112507"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3005985723
Firm city
Plantation
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register