FDA Medical Device Recalls (openFDA)
Zimmer Manufacturing B.V.: Product 29 consists of all product under product code: LPHand same usage: Item no: 32902603800 FEMORAL HEAD +0.5 X 22 MM 32902604810 FEMORAL HEAD +3.5 X 22 MM 32902604845 FEMORAL HEAD +7.0 X 22 MM For use in total or hemi hip arthroplasty
- Recall number
- Z-1127-2016
- Recalling firm
- Zimmer Manufacturing B.V.
- Product
- Product 29 consists of all product under product code: LPHand same usage: Item no: 32902603800 FEMORAL HEAD +0.5 X 22 MM 32902604810 FEMORAL HEAD +3.5 X 22 MM 32902604845 FEMORAL HEAD +7.0 X 22 MM For use in total or hemi hip arthroplasty
- Status
- Terminated
- Initiated
- 2016-01-11
- Posted
- 2016-03-12
- Terminated
- 2017-04-13
- Reason
- As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
- Root cause
- Packaging
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
- Quantity
- 0
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- lot no.: 62051735 62221046 62262799 62272099 62279770 62279770N 62079181 62162466 62185965 62185966 62221047 62262801 62262802 62262803 62279771 62285361 62311660 62393063 62279771N 62285361N 62311660N 61520812 62162464 62221048 62279772 62393064 62279772N 62393064V
- 510(k) numbers
- ["K953337"]
- PMA numbers
- not on file
- Product code
- LPH
- Firm FEI
- 1000132037
- Firm city
- Mercedita
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28