FDA Medical Device Recalls (openFDA)
MEDLINE INDUSTRIES, LP - Northfield: Medline Kits, trays, and packs labeled as follows: a) CATARACT PACK, REF VAL011CTASD; b) CATARACT PACK, REF DYNJ68875F; c) CATARACT PACK, REF DYNJ54869F; d) CATARACT PREP PACK, REF DYNJ44917L; e) E-BLEPH CDS, REF CDS983585G; f) EYE CDS, REF CDS983916G; g) EYE KIT, REF DYKM2222; h) EYE PLASTIC PACK, REF DYNJ47859C; i) K PACK, REF LYN030KAGVD; j) OCULOPLASTY PACK, REF DYNJ64572A; k) OL OCULAR, REF DYNJ906183J; l) OL OCULAR, REF DYNJ906183K; m) P M CATARACT, REF DYNJ44917G; n) P M CATARACT, REF DYNJ44917I; o) P PACK, REF LYN030POGVD; p) R VITRECTOMY, REF EYP087W; q) VITRECTOMY PACK, REF DYNJ68884B
- Recall number
- Z-1126-2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Kits, trays, and packs labeled as follows: a) CATARACT PACK, REF VAL011CTASD; b) CATARACT PACK, REF DYNJ68875F; c) CATARACT PACK, REF DYNJ54869F; d) CATARACT PREP PACK, REF DYNJ44917L; e) E-BLEPH CDS, REF CDS983585G; f) EYE CDS, REF CDS983916G; g) EYE KIT, REF DYKM2222; h) EYE PLASTIC PACK, REF DYNJ47859C; i) K PACK, REF LYN030KAGVD; j) OCULOPLASTY PACK, REF DYNJ64572A; k) OL OCULAR, REF DYNJ906183J; l) OL OCULAR, REF DYNJ906183K; m) P M CATARACT, REF DYNJ44917G; n) P M CATARACT, REF DYNJ44917I; o) P PACK, REF LYN030POGVD; p) R VITRECTOMY, REF EYP087W; q) VITRECTOMY PACK, REF DYNJ68884B
- Status
- Open, Classified
- Initiated
- 2023-11-15
- Posted
- 2024-02-25
- Terminated
- not on file
- Reason
- Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- Root cause
- Nonconforming Material/Component
- Action
- MEDLINE issued an IMMEDIATE ACTION REQUIRED NURSE ASSIST PRODUCT RECALL notice to its consignees on 11/16/2023 via USPS first class mail and email. The notice explained the problem, requested log on to their portal to identify the products and check their inventory so the product could be quarantined. Once Medline receives the completed response form, over-labels will be shipped to the consignee. Medline Industries, LP instructed customers to over-label the convenience kits with a warning label indicating to remove the sodium chloride irrigation or Sterile water and replace from a different sterile supply. Distributors or consignees who further distributed the affected products were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Quantity
- 5724 units
- Distribution
- Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama
- Lot, serial or model codes
- a) REF VAL011CTASD, UDI/DI 40195327152032 (case), 10195327152031 (unit), Lot Numbers: 22LBB812, 22LBH455; b) REF DYNJ68875F, UDI/DI 40195327382125 (case), 10195327382124 (unit), Lot Numbers: 23FBQ273, 23GBS122, 23GBS486, 23HBP752; c) REF DYNJ54869F, UDI/DI 40195327407286 (case), 10195327407285 (unit), Lot Numbers: 23FBQ273, 23GBS122, 23GBS486, 23HBP752; d) REF DYNJ44917L, UDI/DI 40195327472604 (case), 10195327472603 (unit), Lot Numbers: 23HBS811; e) REF CDS983585G, UDI/DI 40193489460781 (case), 10193489460780 (unit), Lot Numbers: 22DBI589, 22HBA111, 22JBQ351, 23EBG770; f) REF CDS983916G, UDI/DI 40193489803861 (case), 10193489803860 (unit), Lot Numbers: 22CBV428, 22GBP859, 22HBP383, 23EBP259, 23GBG620, 23IBS207, 23JBW354; g) REF DYKM2222, UDI/DI 40195327337255 (case), 10195327337254 (unit), Lot Numbers: 23DBI071; h) REF DYNJ47859C, UDI/DI 40193489916066 (case), 10193489916065 (unit), Lot Numbers: 23EBA347; i) REF LYN030KAGVD, UDI/DI 40193489956574 (case), 10193489956573 (unit), Lot Numbers: 22ABH632, 22ABW357, 22CBL660, 22IBS624; j) REF DYNJ64572A, UDI/DI 40195327112555 (case), 10195327112554 (unit), Lot Numbers: 22GBL794; k) REF DYNJ906183J, UDI/DI 40195327279449 (case), 10195327279448 (unit), Lot Numbers: 22LMC016, 22LMH698, 23AME088, 23AMF669, 23BMH365, 23CMB182, 23EMA555; l) REF DYNJ906183K, UDI/DI 40195327385522 (case), 10195327385521 (unit), Lot Numbers: 23FMC104, 23FME336, 23GMD288, 23HMH680; m) REF DYNJ44917G, UDI/DI 40195327031481 (case), 10195327031480 (unit), Lot Numbers: 22BBV644; n) REF DYNJ44917I, UDI/DI 40195327156375 (case), 10195327156374 (unit), Lot Numbers: 22FBN595, 22IBK002; o) REF LYN030POGVD, UDI/DI 22ABM649, 22ABW358, 22BBK347, 22CBT855, 22EBR210, 22HBR247, 22KBH816, 22LBA956, 22OBF733, 23ABB019, 23BBQ119; p) REF EYP087W, UDI/DI 50653160050221 (case), 00653160050226 (unit), Lot Numbers: 40195327037254 (case), 10195327037253 (unit), Lot Numbers: 22CBF187, 22EBE045, 22FBO783, 22GBZ358, 22IBF001, 22JBC220, 22KBN864, 22OBE478, 23ABG110, 23BBR173, 23DBH202; q) REF DYNJ68884B, UDI/DI 40195327059010 (case), 10195327059019 (unit), Lot Numbers: 22GBG399
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJK
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28