TELEFLEX MEDICAL INC: (1) Gibeck Humid-Vent 2S Flex DC, Product #17731, sterile; (2) Gibeck Humid-Vent 2S Flex DC, Product #17732, non-sterile; (3) Gibeck Humid-Vent Filter Compact S with Smooth-Flo/Swivel, Product #19061, sterile; (4) Gibeck Humid-Vent Filter Compact S with Smooth-Flo/Swivel, Product #19062, non-sterile; and (5) Gibeck Humid-Vent Filter Compact S with Flex Tube, Product #19812, non-sterile.
- Recall number
- Z-1126-2022
- Recalling firm
- TELEFLEX MEDICAL INC
- Product
- (1) Gibeck Humid-Vent 2S Flex DC, Product #17731, sterile; (2) Gibeck Humid-Vent 2S Flex DC, Product #17732, non-sterile; (3) Gibeck Humid-Vent Filter Compact S with Smooth-Flo/Swivel, Product #19061, sterile; (4) Gibeck Humid-Vent Filter Compact S with Smooth-Flo/Swivel, Product #19062, non-sterile; and (5) Gibeck Humid-Vent Filter Compact S with Flex Tube, Product #19812, non-sterile.
- Status
- Open, Classified
- Initiated
- 2022-04-11
- Posted
- not on file
- Terminated
- not on file
- Reason
- Reports received of torn corrugated breathing tubes.
- Root cause
- Device Design
- Action
- The recalling firm issued letters dated April 2022 on 4/11/2022 via FedEx 2-day mail. The letter described the issue, the immediate actions required including the hazards involved, and provided specific actions to be taken and instructions to be followed by the medical facilities or the distributors. For the medical facilities, they are to cease use and distribution of the impacted product and immediately quarantine it. If the product is in inventory, indicate the information on the Acknowledgment Form and FAX it to Customer Service or email it to the firm. A customer service representative will contact the medical facility with a Return Goods Authorization number and provide instructions for return of the product to Teleflex. The Acknowledgment Form should also be returned if no product is in inventory. For distributors, they are to provide the recall letter to all customers who have received the product. Each customer is required to complete and return the Acknowledgement Form. The distributor is also supposed to check their inventory, cease distribution and quarantine the affected product. The Acknowledgment Form is to be completed and returned for product in their inventory as well as if they have no inventory. If the distributor has distributed product outside of the country, they are to notify Teleflex Customer Service. Additionally, the recall letter says that this notice should be passed onto all persons who need to be aware within their organization or to any organizations where the potentially affected devices have been transferred. This would include users, clinicians, risk managers, supply chain/distribution centers, etc. in the circulation of the notice.
- Quantity
- 59,581 eaches
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The countries of Austria, Australia, Belgium, Canada, Chile, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Korea, Malaysia, Netherlands, Panama, Poland, Slovakia, Spain, and Switzerland.
- Lot, serial or model codes
- Batch and UDI numbers: (1) Product 17731: KMH21C0130 (01)04026704347919(17)260228(10)KMH21C0130; KMH21C0162 (01)04026704347919(17)260228(10)KMH21C0162; KMH21C0262 (01)04026704347919(17)260228(10)KMH21C0262; KMH21C0265 (01)04026704347919(17)260228(10)KMH21C0265; KMH21E0077 (01)04026704347919(17)260428(10)KMH21E0077; KMH21F0231 (01)04026704347919(17)260528(10)KMH21F0231; KMH21F0276 (01)04026704347919(17)260528(10)KMH21F0276; KMH21G0428 (01)04026704347919(17)260628(10)KMH21G0428; KMH21H0142 (01)04026704347919(17)260728(10)KMH21H0142; KMH21L0020 (01)04026704347919(17)261028(10)KMH21L0020; and KMH22A0068 (01)04026704347919(17)261228(10)KMH22A0068. (2) Product 17732: KMZ21G0733 (01)04026704347926(17)260628(10)KMZ21G0733; KMZ21G0742 (01)04026704347926(17)260628(10)KMZ21G0742; KMZ21G0989 (01)04026704347926(17)260628(10)KMZ21G0989; KMZ21G1049 (01)04026704347926(17)260628(10)KMZ21G1049; KMZ21K0159 (01)04026704347926(17)260928(10)KMZ21K0159; KMZ21K0160 (01)04026704347926(17)260928(10)KMZ21K0160; KMZ21K0275 (01)04026704347926(17)260928(10)KMZ21K0275; KMZ21K0451 (01)04026704347926(17)260928(10)KMZ21K0451; KMZ21L0103 (01)04026704347926(17)261028(10)KMZ21L0103; KMZ21L0602 (01)04026704347926(17)261028(10)KMZ21L0602; KMZ21C0125 (01)04026704347926(17)260228(10)KMZ21C0125; KMZ21C0235 (01)04026704347926(17)260228(10)KMZ21C0235; KMZ21C0300 (01)04026704347926(17)260228(10)KMZ21C0300; KMZ21C0347 (01)04026704347926(17)260228(10)KMZ21C0347; KMZ21C0351 (01)04026704347926(17)260228(10)KMZ21C0351; KMZ21C0475 (01)04026704347926(17)260228(10)KMZ21C0475; KMZ21C0609 (01)04026704347926(17)260228(10)KMZ21C0609; KMZ21C0851 (01)04026704347926(17)260228(10)KMZ21C0851; KMZ21C0863 (01)04026704347926(17)260228(10)KMZ21C0863; KMZ21C1154 (01)04026704347926(17)260228(10)KMZ21C1154; KMZ21C1342 (01)04026704347926(17)260228(10)KMZ21C1342; KMZ21D0016 (01)04026704347926(17)260328(10)KMZ21D0016; KMZ21D0200 (01)04026704347926(17)260328(10)KMZ21D0200; KMZ21D0311 (01)04026704347926(17)260328(10)KMZ21D0311; KMZ21D0357 (01)04026704347926(17)260328(10)KMZ21D0357; KMZ21D0404 (01)04026704347926(17)260328(10)KMZ21D0404; KMZ21D0481 (01)04026704347926(17)260328(10)KMZ21D0481; KMZ21D0510 (01)04026704347926(17)260328(10)KMZ21D0510; KMZ21E0096 (01)04026704347926(17)260428(10)KMZ21E0096; KMZ21F0475 (01)04026704347926(17)260528(10)KMZ21F0475; KMZ21F0483 (01)04026704347926(17)260528(10)KMZ21F0483; KMZ21F0484 (01)04026704347926(17)260528(10)KMZ21F0484; and KMZ21F0873 (01)04026704347926(17)260528(10)KMZ21F0873. (3) Product 19061: KMH21F0418 (01)04026704388387(17)260528(10)KMH21F0418; KMH21G0357 (01)04026704388387(17)260628(10)KMH21G0357; KMH21H0060 (01)04026704388387(17)260728(10)KMH21H0060; KMH21H0062 (01)04026704388387(17)260728(10)KMH21H0062; KMH21H0252 (01)04026704388387(17)260728(10)KMH21H0252; KMH21J0021 (01)04026704388387(17)260828(10)KMH21J0021; KMH21J0260 (01)04026704388387(17)260828(10)KMH21J0260; KMH21K0159 (01)04026704388387(17)260928(10)KMH21K0159; KMH21K0375 (01)04026704388387(17)260928(10)KMH21K0375; KMH21L0139 (01)04026704388387(17)261028(10)KMH21L0139; KMH21C0188 (01)04026704388387(17)260228(10)KMH21C0188; KMH21C0191 (01)04026704388387(17)260228(10)KMH21C0191; KMH21C0194 (01)04026704388387(17)260228(10)KMH21C0194; KMH21C0280 (01)04026704388387(17)260228(10)KMH21C0280; KMH21C0388 (01)04026704388387(17)260228(10)KMH21C0388; KMH21D0094 (01)04026704388387(17)260328(10)KMH21D0094; KMH21D0289 (01)04026704388387(17)260328(10)KMH21D0289; KMH21D0290 (01)04026704388387(17)260328(10)KMH21D0290; KMH21E0240 (01)04026704388387(17)260428(10)KMH21E0240; and KMH21F0155 (01)04026704388387(17)260528(10)KMH21F0155. (4) Product 19062: KMZ21C0425 (01)04026704388394(17)260228(10)KMZ21C0425; KMZ21C0923 (01)04026704388394(17)260228(10)KMZ21C0923; and KMZ21F0379 (01)04026704388394(17)260528(10)KMZ21F0379. (5) Product 19812: KMZ21G0681 (01)04026704388721(17)260628(10)KMZ21G0681; KMZ21G0710 (01)04026704388721(17)260628(10)KMZ21G0710; KMZ21K0459 (01)04026704388721(17)260928(10)KMZ21K0459; KMZ21C0141 (01)04026704388721(17)260228(10)KMZ21C0141; KMZ21C0948 (01)04026704388721(17)260228(10)KMZ21C0948; KMZ21F0361 (01)04026704388721(17)260528(10)KMZ21F0361; and KMZ21F0627 (01)04026704388721(17)260528(10)KMZ21F0627.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- BYD
- Firm FEI
- 3005747797
- Firm city
- Morrisville
- Firm state
- NC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28