Confirm Biosciences Inc: Orient Gene COVID- 19 IgG/ IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) REF GCCOV-402a - Product Usage: is a lateral flow immunoassay intended for the qualitative detection and differentiation of IgM and IgG antibodies to SARS-CoV- 2 in human venous whole blood, plasma from anticoagulated blood or serum.

Recall number
Z-1126-2021
Recalling firm
Confirm Biosciences Inc
Product
Orient Gene COVID- 19 IgG/ IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) REF GCCOV-402a - Product Usage: is a lateral flow immunoassay intended for the qualitative detection and differentiation of IgM and IgG antibodies to SARS-CoV- 2 in human venous whole blood, plasma from anticoagulated blood or serum.
Status
Terminated
Initiated
2021-01-29
Posted
not on file
Terminated
2022-12-13
Reason
Due to two (2) issues: 1) The firm distributed COVID-19 Rapid Test Cassettes outside of the Emergency Use Authorization to non-certified Clinical Laboratories. 2) COVID-19 Rapid Test Cassettes were not adequately validated for its intended use (fingerstick tests for in-home use/testing).
Root cause
Other
Action
On 01/29/2021, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Notification its customers and distributors informing them that Recalling Firm may have distributed products to facilities that do not meeting the conditions outline in the Emergency Use Authorization (EUA) which stipulates that only laboratories certified under Clinical Laboratory Improvement Amendments of 1988 (CLIA) to perform moderate or high complexity tests are authorized to receive and utilize the products. Use of these tests by non-certified laboratory facilities may result in false-positive test results and may lead to an incorrect assessment, such as mistaken immune response to the COVID-19 virus and may delay treatment and risk the further spread of infection to others. The Recalling Firm is planning on sending a follow-up Customer Notification Letter via UPS on or about 02/12/2021. The "URGENT: MEDICAL DEVICE RECALL" Notification instructs customers: 1) To minimize unauthorized use of these products, discontinue the use these products outside of a CLIA accredited or compliant Moderate or High Complexity laboratory setting. 2) Acknowledge the receipt of this notice by completing the attached Response Form, scanning and emailing the completed form to: Regulatory@confirmbiosciences.com Customer are asked to complete the response form and return immediately, no later than March 31, 2021. To correct this issue, the Recalling Firm will work with its customers to either: -Affirm the Customers authorization to use these products through their CLIA Number and Status -Coordinate the return of unused, unopened products to the Recalling Firm. For any questions regarding this recall, call 800-908-5602 or email Regulatory@confirmbiosciences.com (Monday thru Friday 8:00 a.m. to 4:00 p.m. Pacific Time)
Quantity
2,003,053 cassettes
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY. The countries of Australia, Mexico, Honduras, Spain, Sierra Leone, Bahamas, Ecuador, Turkey, Romania, Kuwait, Libya, Qatar, Ghana, Argentina, Bolivia, Brazil, Guatemala, Panama, India, United Kingdom, West Indies, South Africa, United Arab Emirates, Slovenia, Canada, Israel, United Kingdom, Romania, Argentina, Honduras, Russian Federation, R¿publique D¿mocratique du Congo, Switzerland, Dominican Republic, Greece, Saudi Arabia, and Portugal.
Lot, serial or model codes
Catalog Number: HE Cassette COVID 19 Lot Numbers: 2002198, 2003244, 2003285, 2003290, 2003291, 2003295, 2003308, 2004148, 2004158, 2005153, 2005154, 2005155, 2005213
510(k) numbers
not on file
PMA numbers
not on file
Product code
QKO
Firm FEI
3006048451
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register