Flowonix Medical Inc: Prometra II Programmable 20mL Pump. Catalog No. 93827 The device is an implantable infusion pump. GTIN-14 00810335020YYY (YYY= in-country language specific)

Recall number
Z-1125-2020
Recalling firm
Flowonix Medical Inc
Product
Prometra II Programmable 20mL Pump. Catalog No. 93827 The device is an implantable infusion pump. GTIN-14 00810335020YYY (YYY= in-country language specific)
Status
Open, Classified
Initiated
2019-12-24
Posted
2020-02-10
Terminated
not on file
Reason
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
Root cause
Software change control
Action
***Updated 5/12/2023*** On May 11, 2023, the firm sent an URGENT MEDICAL DEVICE CORRECTION UPDATE letter to all affected customers, which informed them of an update to Flowonix's plan to correct the firmware issue. Flowonix will NOT be conducting an office-based pump firmware update for affected pumps implanted prior to January 2020. Healthcare providers were provided ongoing recommendations for Prometra and Prometra II pumps implanted prior to January 2020, which are impacted by this issue. Healthcare providers should contact their Flowonix Representative or our Technical Solutions Department (1-855-356-9665) in the event cessation of therapy or premature low battery is suspected and/or have general questions regarding this notice. *** On December 24, 2019, the firm notified affected users via an Urgent Medical Device Corrective Action Recall Notice. The firm advised customers of the product issue and reminded customers that the pumps are currently labeled only for use with Infumorph. Due to the low risk that this failure mode may occur, Flowonix is not recommending that pumps be explanted. Healthcare providers were recommended to inform patients regarding the symptoms and adverse health events associated with an unexpected pump stoppage and the appropriate steps to take in the event they suspect a stoppage. The firm will be implementing a correction of firmware errors to bring the pumps back into specification. The firm will follow up when the change has been completed. An office-based pump firmware update is being developed for patients currently implanted with the affected pumps. (Note, as of the 5/12/23 update, implanted pumps will not be receiving a firmware update.)
Quantity
1083 (OUS)
Distribution
Nationwide domestic distribution. Foreign distribution to United Kingdom, Republic of Ireland, Poland, Greece, Italy, Germany, Netherlands, and Belgium.
Lot, serial or model codes
All units with pump firmware version 0.26 manufactured from 2012 until the present time.
510(k) numbers
not on file
PMA numbers
["P080012"]
Product code
LKK
Firm FEI
3010079947
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register