Skip to the content

FDA Medical Device Recalls (openFDA)

Medical Components, Inc dba MedComp: TWO PART ADAPTER, REF MCTLA2000, UDI 884908027310 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.

Recall number
Z-1125-2018
Recalling firm
Medical Components, Inc dba MedComp
Product
TWO PART ADAPTER, REF MCTLA2000, UDI 884908027310 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Status
Terminated
Initiated
2018-02-23
Posted
not on file
Terminated
2020-01-10
Reason
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Root cause
Nonconforming Material/Component
Action
On February 23, 2018, Medcomp distributed Product Recall and Distribution notices to international distributors via email and to U.S. facilities via courier service on February 26, 2018. Medcomp is requesting the return of all un-used affected product. Customers should immediately examine inventory and quarantine product subject to recall. If for any reason the product was further distributed, please identify the recipients and notify them at once of this product recall. Your notification may be enhanced by including a copy of this recall notification letter. Contact your customer service representative for a Returned Goods Authorization number if necessary by calling (215) 256-4201. Complete and return the Distribution Notice via fax (215-256-9191) or email (complaints@medcompnet.com.
Quantity
35 kits
Distribution
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
Lot, serial or model codes
Lot Numbers: MCFF700, MCFB160, MKAG11O, MCDP590, MCDS140 & MCFSOSO
510(k) numbers
["K022570"]
PMA numbers
not on file
Product code
NFK
Firm FEI
2518902
Firm city
Harleysville
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register