Covidien LLC: Covidien Nellcor SpO2 Forehead Sensor Item Code: RS10; The Covidien Nellcor Forehead SpO2Sensor, model RS10, is indicated for use when continuous non-invasive arterial oxygen saturation and pulse rate monitoring are required for patients weighing more than 40 kg.

Recall number
Z-1125-2017
Recalling firm
Covidien LLC
Product
Covidien Nellcor SpO2 Forehead Sensor Item Code: RS10; The Covidien Nellcor Forehead SpO2Sensor, model RS10, is indicated for use when continuous non-invasive arterial oxygen saturation and pulse rate monitoring are required for patients weighing more than 40 kg.
Status
Terminated
Initiated
2016-12-16
Posted
not on file
Terminated
2018-12-04
Reason
Labeling error: The label indicates the product is latex free, but the headband component included with the forehead sensor does contain latex.
Root cause
Process control
Action
Medtronic initiated on December 16, 2016. Customers were notified via Federal Express and the letter instructed customers to discontinue use of the product and return all inventory to Medtronic (Covidien) 195 McDermott Road, North Haven, CT 06473 Attn: Field Returns Department. Complete a Recalled Product Return Form (Attachment A, included with letter) and submit it to customer service so a Returned Goods Authorization number (RGA#) can be issued. Questions contact Medtronic representative or customer service at 800-882-5878.
Quantity
7453 Units
Distribution
Nationwide. Foreign: Belgium Germany France Netherlands
Lot, serial or model codes
Lot Numbers: 161510095H, 161620089H, 161690203H, 161760132H, 161900091H,161970090H, 162040107H, 162110103H, 162320191H, 162390212H
510(k) numbers
["K904039"]
PMA numbers
not on file
Product code
DQA
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register