Skip to the content

FDA Medical Device Recalls (openFDA)

CareFusion 303, Inc.: BD Pyxis MedStation Auxiliary 4000 REF: 306. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.

Recall number
Z-1124-2025
Recalling firm
CareFusion 303, Inc.
Product
BD Pyxis MedStation Auxiliary 4000 REF: 306. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.
Status
Open, Classified
Initiated
2024-02-23
Posted
2025-02-11
Terminated
not on file
Reason
Retroactive-Due to increase in complaints, their is a potential for drawer and/or door failures of the medication cabinet which could result in delay in accessing medications.
Root cause
Device Design
Action
On February 14, 2023 and January 9. 2024 Carefusion issued a Urgent medical Device Product Advisory" Recall Notification to affected consignees. Carefusion asked consignees take the following actions: 1. Customers may continue to use these devices as intended and should file a complaint via the Customer Portal at https://www.bd.com/self-service or call BD Support at 1-800-727-6102, should you experience drawer or door failures. 2. If you experience delays in access to medication due to a drawer or door failure, ensure backup procedures (which may include but may not necessarily be limited to keys, emergency medication supplies, alternate devices in proximity, etc.) are in place and staff have all been properly trained on them. 3. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter. 4. If you have specific questions about this letter, please contact 1-866-583-8783. 5. Any adverse health consequences experienced with the use of this product should be reported to BD and may be reported to the FDA s MedWatch Adverse Event Reporting Actions to be taken by distributor 1. Please provide a copy of this notification to any customers who may have received products listed in the Affected Products table above. 2. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter.
Quantity
not on file
Distribution
US: Nationwide OUS:JP IT IN DE GB KR NL TW ES BE AU BH CA BR NZ FR AR QA MX CL TR SA MY CH IL IE TH SG KW MA FI AE EG BS BM GR MC PL LB MO IQ TN NO PH CN HK CO JO OM
Lot, serial or model codes
UDI: 10885403512636/ Serial Numbers: 16187325, 16187326, 16187327, 16173446
510(k) numbers
not on file
PMA numbers
not on file
Product code
BRY
Firm FEI
2016493
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register