Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet U.K., Ltd.: REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants.

Recall number
Z-1124-2013
Recalling firm
Biomet U.K., Ltd.
Product
REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants.
Status
Terminated
Initiated
2013-03-20
Posted
2013-04-15
Terminated
2015-05-04
Reason
Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle.
Root cause
Device Design
Action
BIOMET sent an Urgent Medical Device Recall Notice on March 20, 2013, to all affected customers with directions to forward the notice if product was further distributed to Hospital Operating Room managers. The notice identified the recalled device, reason for the recall, and possible adverse events directing consignees to locate and discontinue use of the device. A FAX back Response form was to be returned to 574-372-1683 within 3 days of receipt prior to return of the product to Biomet, USA. Questionsregarding this recall should be addressed to 574 372-1570 M-F, 8a.m. -5 p.m.
Quantity
430
Distribution
Worldwide Distribution - USA (nationwide), Europe, Japan, Australia, Canada, Singapore, South Africa, and Costa Rica.
Lot, serial or model codes
32-422760 ( sold as part of instrument set 32-422763 ) ZB100901,ZB110201,ZB110301,ZB110302,ZB110303, ZB110304,ZB110601,ZB110602,ZB110701,ZB111201, ZB111202,ZB120201,ZB120202,ZB120701,ZB120801, ZB120802,ZB121102,ZB121103
510(k) numbers
not on file
PMA numbers
not on file
Product code
NRA
Firm FEI
3002806535
Firm city
South Wales
Firm state
Bridgend
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register