FDA Medical Device Recalls (openFDA)
Biomet U.K., Ltd.: REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants.
- Recall number
- Z-1124-2013
- Recalling firm
- Biomet U.K., Ltd.
- Product
- REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants.
- Status
- Terminated
- Initiated
- 2013-03-20
- Posted
- 2013-04-15
- Terminated
- 2015-05-04
- Reason
- Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle.
- Root cause
- Device Design
- Action
- BIOMET sent an Urgent Medical Device Recall Notice on March 20, 2013, to all affected customers with directions to forward the notice if product was further distributed to Hospital Operating Room managers. The notice identified the recalled device, reason for the recall, and possible adverse events directing consignees to locate and discontinue use of the device. A FAX back Response form was to be returned to 574-372-1683 within 3 days of receipt prior to return of the product to Biomet, USA. Questionsregarding this recall should be addressed to 574 372-1570 M-F, 8a.m. -5 p.m.
- Quantity
- 430
- Distribution
- Worldwide Distribution - USA (nationwide), Europe, Japan, Australia, Canada, Singapore, South Africa, and Costa Rica.
- Lot, serial or model codes
- 32-422760 ( sold as part of instrument set 32-422763 ) ZB100901,ZB110201,ZB110301,ZB110302,ZB110303, ZB110304,ZB110601,ZB110602,ZB110701,ZB111201, ZB111202,ZB120201,ZB120202,ZB120701,ZB120801, ZB120802,ZB121102,ZB121103
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NRA
- Firm FEI
- 3002806535
- Firm city
- South Wales
- Firm state
- Bridgend
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28