FDA Medical Device Recalls (openFDA)

Physio-Control, Inc.: LIFEPAK 15 Monitor, Automated External Defibrillator (AED), Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.

Recall number
Z-1123-2017
Recalling firm
Physio-Control, Inc.
Product
LIFEPAK 15 Monitor, Automated External Defibrillator (AED), Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.
Status
Terminated
Initiated
2017-01-12
Posted
2017-02-11
Terminated
2017-09-01
Reason
The firm has become aware of a potential device issue where the LIFEPAK 15 device may no longer be able to deliver defibrillation energy due to a potential failure of an internal component (Relay) installed on the Therapy Printed Circuit Board Assembly (PCBA).
Root cause
Other
Action
Physio plans to send the URGENT MEDICAL DEVICE CORRECTION ACTION REQUIRED LIFEPAK 15 Monitor/Defibrillator letter, dated January 2017, to their consignees on January 12, 2017. They were informed of the problem and how to identify a malfunction during the User Test of the device as described in the chapter titled Maintaining the Equipment within the LIFEPAK 15 Operating Instructions (or follow page 2 of the notification letter). If the User Test fails, the device will illuminate the Service indicator on the device. Please contact Physio-Control immediately to schedule the device correction. This field corrective action will include replacement of the Therapy PCB assembly on all devices included in the potentially affected population. This will occur as a combination of field and depot service. Customers are required to take the following actions: 1. Please forward this letter to all of your sites, trainers and users that have an affected LIFEPAK 15 device as identified on the attached Confirmation Sheet. 2. Follow the instructions on the Confirmation Sheet for each serial number listed in your possession. Promptly return the completed Confirmation Sheet to Physio-Control. 3. If the device fails the User Test, as described above, contact Physio-Control immediately to arrange for correction of your device. 4. Continue to perform the User Test as outlined in the Daily Operators Checklist within the LIFEPAK 15 Operating Instructions. Consignees should return completed confirmation sheet to Physio: -By fax to: 1-866-448-9567 -By email to: rsrecall@physio-control.com -Or by mail to: Physio-Control, Inc. P.O. Box 97006, Dept. 15N Redmond, WA, 98073-9706. Customers with questions call the firm at 1-800-442-1142, option 7, 6:00 A.M. to 4:00 P.M. (PST), Monday Friday.
Quantity
338 units total (US =139 units; International = 199 units)
Distribution
Worldwide Distribution-US ( nationwide) in states of: AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TX, VA, WA, WI, and WV and countries of: Australia, Brazil, Canada, Czech Republic, Denmark, Finland, France, French Guiana, Germany, India, Ireland, Israel, Italy, Japan, Latvia, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Qatar, Saudi Arabia, Slovenia, South Africa, Sweden, Turkey, United Arab Emirates, and United Kingdom.
Lot, serial or model codes
SERIAL NUMBERS:  42738872, 44554102, 42701550, 42702539, 42730815, 42800549, 42803986, 42862109, 42884428, 42944609, 42989327, 43022460, 43040692, 43048791, 43066974, 43116838, 43116994, 43119477, 43132954, 43133203, 43133221, 43133906, 43154308, 43154457, 43156695, 43161425, 43163930, 43164076, 43164158, 43166133, 43206034, 43210019, 43219622, 43226291, 43230032, 43232487, 43232904, 43247389, 43251050, 43253506, 43253916, 43254032, 43258901, 43293290, 43321556, 43388132, 43388137, 43388541, 43405542, 43407947, 43409346, 43429508, 43465372, 43472036, 43472591, 43480824, 43489606, 43492883, 43576021, 43616526, 43641893, 43691775, 43715563, 43735849, 43764249, 43800357, 43805213, 43825370, 43826635, 43843347, 43869096, 43884778, 43922984, 43938506, 43947198, 43962139, 43964195, 43968233, 43981293, 43999541, 44010117, 44010465, 44039473, 44056838, 44060132, 44072885, 44073149, 44090669, 44092118, 44094122, 44108545, 44112183, 44117966, 44150897, 44167052, 44168042, 44179750, 44199424, 44225054, 44225258, 44226227, 44226585, 44226654, 44231094, 44233618, 44234803, 44234840, 44237875, 44255648, 44262283, 44262826, 44264102, 44264573, 44265421, 44271391, 44283558, 44287000, 44289115, 44293297, 44303302, 44312782, 44315479, 44327764, 44342010, 44352664, 44372343, 44416667, 44435825, 44438567, 44439898, 44449825, 44450073, 44452220, 44458445, 44460124, 44460749, 44463853, 44466514, 44473363, 44483809, 44486872, 44487784, 44487992, 44488101, 44489681, 44492189, 44492223, 44492252, 44492760, 44492832, 44494129, 44496734, 44497169, 44497196, 44497761, 44497866, 44498576, 44500827, 44500943, 44501037, 44501047, 44501112, 44501201, 44501465, 44501806, 44501832, 44502009, 44502159, 44502179, 44502310, 44502686, 44502718, 44502776, 44502815, 44502943, 44503096, 44508212, 44508273, 44508330, 44508331, 44508358, 44508440, 44508498, 44508645, 44508868, 44509074, 44509520, 44509538, 44511211, 44511231, 44511273, 44511519, 44511573, 44512604, 44513321, 44513358, 44518244, 44518479, 44518497, 44518703, 44518786, 44518917, 44518981, 44519083, 44519095, 44519255, 44519301, 44519307, 44519381, 44519548, 44524967, 44525130, 44525416, 44527324, 44530479, 44530594, 44530719, 44530800, 44531488, 44531803, 44532088, 44534804, 44534900, 44534991, 44535111, 44535241, 44535452, 44535582, 44535643, 44536176, 44536780, 44536823, 44536897, 44538798, 44538819, 44539781, 44539880, 44541779, 44541917, 44548263, 44548530, 44548558, 44548749, 44548825, 44549064, 44549220, 44549413, 44549964, 44550150, 44550234, 44550467, 44550552, 44550575, 44550613, 44550642, 44550675, 44552029, 44552066, 44552865, 44553122, 44554175, 44554259, 44554351, 44554520, 44554837, 44554941, 44555058, 44555261, 44556688, 44557824, 44558175, 44558252, 44558385, 44559077, 44559369, 44559455, 44560001, 44560185, 44561655, 44561903, 44562074, 44562559, 44562662, 44562754, 44563762, 44564474, 44564491, 44571952, 44572025, 44572171, 44572228, 44572357, 44572374, 44572595, 44572865, 44573101, 44573148, 44573485, 44573956, 44574467, 44574766, 44574771, 44574821, 44574929, 44574939, 44575793, 44575871, 44575898, 44576359, 44576647, 44577004, 44577022, 44577116, 44577219, 44578054, 44578080, 44578192, 44578272, 44579036, 44579299, 44579496, 44579697, 44579713, 44579857, 44581304, 44581382, 44583260, 44583657, 44583701, 44584213, 44586462, 44587097, 44587099, 44587181, 44587302, 44587358, 44587899, and 44587976.
510(k) numbers
["K142430"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
3015876
Firm city
Redmond
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register