FDA Medical Device Recalls (openFDA)
Mazor Robotics Ltd: Mazor X robotic guidance system REF: TPL0059
- Recall number
- Z-1122-2026
- Recalling firm
- Mazor Robotics Ltd
- Product
- Mazor X robotic guidance system REF: TPL0059
- Status
- Open, Classified
- Initiated
- 2025-12-10
- Posted
- 2026-01-16
- Terminated
- not on file
- Reason
- Software errors that can result in incorrect surgical instrument positioning during spinal surgery.
- Root cause
- Under Investigation by firm
- Action
- Beginning December 10, 2025, Medtronic (Mazor Robotics, Ltd) issued a Urgent Medical Device Recall Notification to affected consignees via E-Mail. Medtronic asked consignees to take the following actions: 1. Please review this information with all physician users and/or post a copy of this notification with your Mazor X system until the software update is completed. 2. Please confirm via the enclosed confirmation form that this notification has been communicated within your facility with all physician users. Send the completed Customer Confirmation Form to Medtronic via email at neuro.quality@medtronic.com. 3. Provide this notification to those who need to be aware within your organization or to any organization where the affected systems have been transferred. 4. Maintain a copy of this notice in your records. 5. Continue to use the Mazor X system as instructed in the user manual and with the mitigations provided below in Appendix A.
- Quantity
- 549 systems
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, China, Croatia, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States, Vietnam.
- Lot, serial or model codes
- Software versions: 5.0.1, 5.1.2, 5.1.3/ UDI: 07290109180465, 07290109184524, 07290109181158, 00763000635169, 07290109183213, 07290109184517, 00763000419134, 00763000431761, 07290109184098, 07290109184838, 07290115751376, 07290115751895
- 510(k) numbers
- ["K230064"]
- PMA numbers
- not on file
- Product code
- OLO
- Firm FEI
- 3005075696
- Firm city
- Caesarea
- Firm state
- not on file
- Firm country
- Israel
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28