FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corporation: Gambro Cartridge Low Volume Blood Transport System for Hemodialysis Blood Set, product code 114434 - Product Usage: intended for single use in a hemodialysis treatment using the Phoenix Dialysis Delivery Systems.
- Recall number
- Z-1122-2021
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Gambro Cartridge Low Volume Blood Transport System for Hemodialysis Blood Set, product code 114434 - Product Usage: intended for single use in a hemodialysis treatment using the Phoenix Dialysis Delivery Systems.
- Status
- Open, Classified
- Initiated
- 2020-12-30
- Posted
- not on file
- Terminated
- not on file
- Reason
- complaints of kinked access lines observed during treatment using cartridge sets for hemodialysis treatments were received.
- Root cause
- Under Investigation by firm
- Action
- The firm issued a Safety Alert on 12/30/2020 by mail. The letter explained the problem with photographs, and the hazard involved with using a device with a kinked line. Customers identifying set with kinked tubing may contact Baxter Healthcare Center for Service to arrange for return and credit. Baxter Healthcare Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. Distributor were asked to notify their customers by distributing the letter.
- Quantity
- 57825 devices
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- UDI 07332414112394, All lots within expiry
- 510(k) numbers
- ["K070414"]
- PMA numbers
- not on file
- Product code
- FJK
- Firm FEI
- 1417572
- Firm city
- Deerfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28