Skip to the content

FDA Medical Device Recalls (openFDA)

WalkMed Infusion, LLC: Triton FP Infusion Pump, Part Number 400000, packaged in boxes. The Triton Volumetric Infusion Pump is indicated for delivering infusions of medications, as well as providing TPN (Total Parenteral Nutrition) and PCA (Patient Control Analgesia). It is intended for use in hospitals, ambulatory & nursing home (extended care) settings, and home care environments.

Recall number
Z-1122-2014
Recalling firm
WalkMed Infusion, LLC
Product
Triton FP Infusion Pump, Part Number 400000, packaged in boxes. The Triton Volumetric Infusion Pump is indicated for delivering infusions of medications, as well as providing TPN (Total Parenteral Nutrition) and PCA (Patient Control Analgesia). It is intended for use in hospitals, ambulatory & nursing home (extended care) settings, and home care environments.
Status
Terminated
Initiated
2014-02-17
Posted
2014-03-03
Terminated
2014-10-06
Reason
WalkMed Infusion is initiaing a recall on their Triton FP Infusion Pump, Part Number 400000 due to the label on the right side of the pump containing an incorrect instruction.
Root cause
Labeling design
Action
WalkMed Infusion sent an Important Medical Device Recall letter dated February 14, 2014, to all affected customers. The letter identified the product the problem and the action customers need to take. Our records Indicate that you have purchased product affected by this recall (see the attached list).Our sales representative will contact you to determine a plan for label replacement for your units. Replacement of the label will include a simple process of removing the old label (PN 300063), replacing it with a corrected label(PN 400033) and documenting completion of the replacement. This action can be completed by you,your customers,our sales representative, or by returning the infusion pump to WalkMed Infusion. For further questions please call (720) 351-4934 or (720) 351-4965. Product will be corrected in the field when possible and supported by the customer. Otherwise, WalkMed Infusion employees with travel to the customer and complete the correction activities.
Quantity
797
Distribution
US Distribution including the states of IL, MA, KS, and, NJ.
Lot, serial or model codes
PFA14001,PFA14002,PFA14003,PFA14004,PFA14005,PFA14006,PFA14007,PFA14008,PFA14009,PFA14010,PFA14011,PFA14012,PFA14013,PFA14014,PFA14015,PFA14016,PFA14017,PFA14018,PFA14019,PFA14020,PFA14021,PFA14022,PFA14023,PFA14024,PFA14025,PFA14026,PFA14027,PFA14028,PFA14029,PFA14030,PFA14031,PFA14032,PFA14033,PFA14034,PFA14035,PFA14036,PFA14037,PFA14038,PFA14039,PFA14040,PFA14041,PFA14042,PFA14043,PFA14044,PFA14045,PFA14046,PFA14047,PFA14048,PFA14049,PFA14050,PFA14051,PFA14052,PFA14053,PFA14054,PFA14055,PFA14056,PFA14057,PFA14058,PFA14059,PFA14060,PFK13106,PFK13107,PFK13108,PFK13109,PFK13110,PFK13111,PFK13112,PFK13113,PFK13114,PFK13115,PFK13116,PFK13117,PFK13118,PFK13119,PFK13120,PFK13121,PFK13122,PFK13123,PFK13124,PFK13125,PFK13126,PFK13127,PFK13128,PFK13129,PFK13130,PFK13131,PFK13132,PFK13133,PFK13134,PFK13135,PFK13136,PFK13137,PFK13138,PFK13139,PFK13140,PFK13141,PFK13142,PFK13143,PFK13144,PFK13145,PFL13001,PFL13002,PFL13003,PFL13004,PFL13005,PFL13006,PFL13007,PFL13008,PFL13009,PFL13010,PFL13011,PFL13012,PFL13013,PFL13014,PFL13015,PFL13016,PFL13017,PFL13018,PFL13019,PFL13020,PFL13021,PFL13022,PFL13023,PFL13024,PFL13025,PFL13026,PFL13027,PFL13028,PFL13029,PFL13030,PFL13031,PFL13032,PFL13033,PFL13034,PFL13035,PFL13036,PFL13037,PFL13038,PFL13039,PFL13040,PFL13041,PFL13042,PFL13043,PFL13044,PFL13045,PFL13046,PFL13047,PFL13048,PFL13049,PFL13050,
510(k) numbers
["K070529"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3000204189
Firm city
Englewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register