Skip to the content

FDA Medical Device Recalls (openFDA)

MICROVENTION INC.: MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended purpose of the LVIS Device is to serve as a stent-assisted coiling (SAC) device providing support to the embolic coils used in the treatment of wide-necked aneurysms.

Recall number
Z-1121-2025
Recalling firm
MICROVENTION INC.
Product
MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended purpose of the LVIS Device is to serve as a stent-assisted coiling (SAC) device providing support to the embolic coils used in the treatment of wide-necked aneurysms.
Status
Open, Classified
Initiated
2024-12-11
Posted
2025-02-07
Terminated
not on file
Reason
Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent from being pushed out of the introducer.
Root cause
Nonconforming Material/Component
Action
On December 11, 2024, MicroVention Europe SARL aka Terumo Neuro issued a "Urgent Medical Device Recall" Notification to affected consignees via mail. Terumo Neuro asked consignees to take the following actions: 1. Review Inventory; 2. Inform and Forward Recall Notice; 3. Return Product; 4. Please complete and return the "Customer Acknowledgment Form"; 4. Contact Quality Department ; 5. Replacement or Credit of Returned Devices
Quantity
11 units
Distribution
US: None OUS: China
Lot, serial or model codes
Lot number: 0000456768
510(k) numbers
not on file
PMA numbers
["P170013"]
Product code
QCA
Firm FEI
3013556777
Firm city
Aliso Viejo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register