Accriva Diagnostics, Inc.: The VerifyNow P2Y12 Test is designed to measure platelet P2Y12 receptor blockade. Substances known to specifically block the P2Y12 receptor include ticagrelor and the thienopyridine class of drugs, including clopidogrel and prasugrel.1-8 The test is based upon the ability of activated platelets to bind fibrinogen. Fibrinogen-coated microparticles aggregate in whole blood in proportion to the number of expressed platelet GP IIb/IIIa receptors. Light transmittance increases as activated platelets bind and aggregate fibrinogen-coated beads. The instrument measures this change in optical signal and reports results in P2Y12 Reaction Units (PRU). The rate of microbead aggregation is more rapid and reproducible if platelets are activated; therefore, the reagent adenosine-5-diphosphate (ADP) is incorporated into the test channel to induce platelet activation. Prostaglandin E1 (PGE1) is included in the reagent mixture to increase the specificity of the PRU result to reflect P2Y12-mediated platelet aggregation. The reagents are formulated such to achieve reaction concentrations of 20 M ADP and 22 nM PGE1. In a like manner, two other activators, iso-TRAP (Thrombin Receptor Activating Peptide) with PAR4-AP (PAR4-Activating Peptide) and fibrinogen-coated microparticles are incorporated into a second channel of the test device. The instrument measures the change in transmittance in this channel, calculates the baseline platelet function for the sample, and reports the percent inhibition result for the sample.
- Recall number
- Z-1121-2022
- Recalling firm
- Accriva Diagnostics, Inc.
- Product
- The VerifyNow P2Y12 Test is designed to measure platelet P2Y12 receptor blockade. Substances known to specifically block the P2Y12 receptor include ticagrelor and the thienopyridine class of drugs, including clopidogrel and prasugrel.1-8 The test is based upon the ability of activated platelets to bind fibrinogen. Fibrinogen-coated microparticles aggregate in whole blood in proportion to the number of expressed platelet GP IIb/IIIa receptors. Light transmittance increases as activated platelets bind and aggregate fibrinogen-coated beads. The instrument measures this change in optical signal and reports results in P2Y12 Reaction Units (PRU). The rate of microbead aggregation is more rapid and reproducible if platelets are activated; therefore, the reagent adenosine-5-diphosphate (ADP) is incorporated into the test channel to induce platelet activation. Prostaglandin E1 (PGE1) is included in the reagent mixture to increase the specificity of the PRU result to reflect P2Y12-mediated platelet aggregation. The reagents are formulated such to achieve reaction concentrations of 20 M ADP and 22 nM PGE1. In a like manner, two other activators, iso-TRAP (Thrombin Receptor Activating Peptide) with PAR4-AP (PAR4-Activating Peptide) and fibrinogen-coated microparticles are incorporated into a second channel of the test device. The instrument measures the change in transmittance in this channel, calculates the baseline platelet function for the sample, and reports the percent inhibition result for the sample.
- Status
- Terminated
- Initiated
- 2022-03-30
- Posted
- not on file
- Terminated
- 2024-11-07
- Reason
- Two whole blood Platelet Reactivity tests exist that share the same reagent formulation and are used to obtain P2Y12 Reactivity Unit results. The test without US-FDA market clearance was distributed and it displays BASE results not displayed by the US-FDA market cleared device. A formula is being provided to health care personnel to calculate the percentage of platelet aggregation inhibition.
- Root cause
- Employee error
- Action
- On 03/30/22, recall notices were distributed to customers who were asked to do the following: 1) Forward the notice within their organization and all other organizations where affected devices may have been transferred to. 2) Identify affected devices and do not use. 3) Destroy all affected devices and document the destruction on the Tracking Form. 4) Complete and return the Tracking Form to the recalling firm. Customers with additional questions can contact Customer Service at 800-955-9525 or e-mail customerservice.na@werfen.com
- Quantity
- 11
- Distribution
- US Nationwide distribution in the state of California.
- Lot, serial or model codes
- Lot Number: WL1060, Expiration: 2023-03-01
- 510(k) numbers
- ["K141427"]
- PMA numbers
- not on file
- Product code
- JOZ
- Firm FEI
- 2250033
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28