Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc: ADVIA Centaur Vitamin D Assay (500 tests); Reference Number:10631021. For use in the quantitative determination of Vit D.

Recall number
Z-1120-2013
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
ADVIA Centaur Vitamin D Assay (500 tests); Reference Number:10631021. For use in the quantitative determination of Vit D.
Status
Terminated
Initiated
2013-03-13
Posted
2013-04-15
Terminated
2015-03-05
Reason
Negative bias in patient samples for Vitamin D
Root cause
No Marketing Application
Action
Siemens Healthcare issued Urgent Medical Device Recall was sent to all affected Siemens Healthcare Diagnostics customers in the United States on March 13th, 2013. This notice informs the customer of the recall and the actions that need to be taken. The Urgent Medical Device Recall instructs customers to discontinue the use of ADVIA Centaur VitD Assay Kit Lots (REF 10491994 and 10631021) and ADVIA Centaur Systems VitD Cal (REF 10493589 and 10630911) released prior to the recalibration of the assay and request replacement lots. This notice also directs customers to review the contents of the recall notice with their laboratory or medical director. A confirmation fax-back form is included to ensure that all customers in the United States have been notified and have complied with the mandatory action.
Quantity
4133
Distribution
Nationwide Distribution.
Lot, serial or model codes
Ready Pack Lot Number: 134009 Kit Lot Number/Exp. Date: 88767009 4/28/2013 88787009 4/28/2013  Ready Pack Lot Number: 134010 Kit Lot Number/Exp. Date: 90253010 6/2/2013 91934010 6/2/2013 Ready Pack Lot Number: 134011 Kit Lot Number/Exp. Date: 92396011 8/12/2013 93363011 8/12/2013 95088011 8/12/2013
510(k) numbers
not on file
PMA numbers
not on file
Product code
MRG
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register