Skip to the content

FDA Medical Device Recalls (openFDA)

Arrow International Inc: Arrow Two-Lumen Central Venous Catheterization Kit with Blue FlexTip Catheter Product Code: AH-11802 (OUS) GTN: 00801902026132

Recall number
Z-1118-2019
Recalling firm
Arrow International Inc
Product
Arrow Two-Lumen Central Venous Catheterization Kit with Blue FlexTip Catheter Product Code: AH-11802 (OUS) GTN: 00801902026132
Status
Terminated
Initiated
2019-01-10
Posted
not on file
Terminated
2020-06-04
Reason
The product lidstock contains a labelling error. The lidstock states the incorrect priming volume. The priming volumes printed on the extension lines are correct
Root cause
Error in labeling
Action
Teleflex issued Urgent Medical Device Notification to US consignee notification letter on 10th January 2019 via FedEx 2-day mail. Actions to be taken to remove product from the facilities: A: The affected customer is instructed to take the following actions: 1. Place a copy of this notification with each unit of affected product currently in your inventory. 2. After step (1) is complete, fill out the enclosed Acknowledgement Form
Quantity
None (US)
Distribution
FL, NC
Lot, serial or model codes
Lot Numbers:  71F14E0696 71F14K0658 71F15E0493 71F15L0621 71F14E1541 71F14K1683 71F15F0090 71F15M0787 71F14F1154 71F14L0719 71F15H0574 71F16A0278 71F14H0474 71F14M1375 71F15H1503 71F16B0054 71F14H1215 71F15A0025 71F15J0570 71F16B1306 71F14J0852 71F15A1592 71F15J2098 71F14J1889 71F15D1460 71F15L0153
510(k) numbers
["K862056"]
PMA numbers
not on file
Product code
DQY
Firm FEI
3015859709
Firm city
Reading
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register