Intuitive Surgical, Inc.: HARMONIC ACE Curved Shears Instructions for Use, 8MM, IS2000/IS3000; for use with the da Vinci Surgical System - IS2000 and IS3000. Intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. It is used in conjunction with the da Vinci Surgical System
- Recall number
- Z-1118-2015
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- HARMONIC ACE Curved Shears Instructions for Use, 8MM, IS2000/IS3000; for use with the da Vinci Surgical System - IS2000 and IS3000. Intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. It is used in conjunction with the da Vinci Surgical System
- Status
- Terminated
- Initiated
- 2015-01-21
- Posted
- 2015-02-13
- Terminated
- 2015-06-01
- Reason
- Internal labeling review found that the IFU for the Harmonic Ace incorrectly instructs the user on how to manually open the grips.
- Root cause
- Error in labeling
- Action
- A Field Safety Notice entitled, Urgent Medical Device Correction, and dated, January 21, 2015, were sent to customers. The letter identified the problem and the reason for the recall as well as the risk to health. Customers are asked to ensure that all affected personnel are informed of the notice. Additionally, customers are to discard their current Harmonic Ace IFUs. Customers are to refer to the specified manual, reference guide, and instruction card mentioned in the letter. A copy of the notice is to be retained with the da Vinci Standard, S or Si user manual. Intuitive Surgical will remove the Harmonic ACE IFU from all future shipments of the product. Representatives will be available by phone to answer questions related to the Correction.
- Quantity
- 12,467 - both sizes
- Distribution
- Worldwide Distribution -- US, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, India, Israel, Italy, Japan, Lebanon, Mexico, Monaco, Netherlands, Pakistan, Panama, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, and the United Kingdom.
- Lot, serial or model codes
- PN 400275, 420275, and 420272.
- 510(k) numbers
- ["K093217"]
- PMA numbers
- not on file
- Product code
- NAY
- Firm FEI
- 3001675293
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28