Skip to the content

FDA Medical Device Recalls (openFDA)

OMNIlife science Inc.: Apex Knee System, Tibial Baseplate Size 6 Left, non-porous - KC-2206L The Apex Knee System is intended for use as a primary or revision total knee replacement. This knee replacement system is intended for cemented fixation and single use implantation. This prosthesis may be used for the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; Rheumatoid arthritis; Correction of functional deformity; Revision procedures where other treatments or devices have failed.

Recall number
Z-1117-2018
Recalling firm
OMNIlife science Inc.
Product
Apex Knee System, Tibial Baseplate Size 6 Left, non-porous - KC-2206L The Apex Knee System is intended for use as a primary or revision total knee replacement. This knee replacement system is intended for cemented fixation and single use implantation. This prosthesis may be used for the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; Rheumatoid arthritis; Correction of functional deformity; Revision procedures where other treatments or devices have failed.
Status
Terminated
Initiated
2017-11-20
Posted
not on file
Terminated
2019-06-07
Reason
The Apex Knee System Tibial Baseplate identified has an improper expiration date marked on the external packaging. The improper expiration date (2202-03) is marked on the external packaging only. The correct lot number (2022-03) is present in the internal packaging and sterile batch documentation.
Root cause
Incorrect or no expiration date
Action
OMNIlife science, Inc. sent an Urgent Medical Device Advisory Notice dated November 20, 2017. Actions to be taken by the Customer/User "Immediately quarantine lot 26519 for KC-2206L "Return Product Using RMA# XXXX to OMNI. "This advisory notice only includes LOT 26519. It does not affect other lots. You must acknowledge this notice by returning the attached Acknowledgement and Receipt Form and accompanying by FAX to 508-819-3390 or my email to CS@omnils.com. For further questions, please call (508) 824-2444.
Quantity
32
Distribution
Worldwide Distribution - US Distribution and to the country Spain.
Lot, serial or model codes
Product Code KC-2206L, Lot # 26519
510(k) numbers
["K060192"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3008342029
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register