Zimmer Manufacturing B.V.: Product 19 consists of all product code: JDI and same usage: Item no: 765301101 CEM FEM ST 11X120 NOCOAT 765301201 CEM FEM ST 12X125 NOCOAT 765301301 CEM FEM ST 13X130 NOCOAT 765301401 CEM FEM ST 14X135 NOCOAT 765301501 CEM FEM ST 15X140 NOCOAT 765301601 CEM FEM ST 16X145 NOCOAT 763301300 CEM FEM ST FX 13 X 130 763301600 CEM FEM ST FX 16 X 145 For use in total or hemi hip arthroplasty

Recall number
Z-1117-2016
Recalling firm
Zimmer Manufacturing B.V.
Product
Product 19 consists of all product code: JDI and same usage: Item no: 765301101 CEM FEM ST 11X120 NOCOAT 765301201 CEM FEM ST 12X125 NOCOAT 765301301 CEM FEM ST 13X130 NOCOAT 765301401 CEM FEM ST 14X135 NOCOAT 765301501 CEM FEM ST 15X140 NOCOAT 765301601 CEM FEM ST 16X145 NOCOAT 763301300 CEM FEM ST FX 13 X 130 763301600 CEM FEM ST FX 16 X 145 For use in total or hemi hip arthroplasty
Status
Terminated
Initiated
2016-01-11
Posted
2016-03-12
Terminated
2017-04-13
Reason
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root cause
Packaging
Action
Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
Quantity
0
Distribution
US Nationwide Distribution
Lot, serial or model codes
lot no.: 62222763 62240266 62268561 62284487 62294353 62310467 62320160 62346566 61846909 61859930 62028715 62069012 62114832 62237560 62278694 62284486 62298294 62305946 62310463 62330767 62340043 61864095 62028713 62038326 62086042 62100678 62114831 62215704 62284488 62291943 62305945 62311644 62317919 62320157 62320158 62340044 62363364 61866237 62105184 62215705 62268566 62284489 62294354 62303512 62311645 62330768 62065287 62211913 62268563 62284490 62299091 62320161 62333990 62299091N 61885871 62100679 62278695 62303703 62333992 62069011 62280472
510(k) numbers
not on file
PMA numbers
not on file
Product code
JDI
Firm FEI
1000132037
Firm city
Mercedita
Firm state
not on file
Firm country
PR

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register