Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Dental Inc: GemLock Long Hex Driver, Catalog RHL2.5; Also distributed within the Tapered SwissPlus & SwissPlus Implant Systems Kits, Surgical, Complete, Catalog OPCST; Tapered Screw-Vent Implant System Surgical Kit, Complete, Catalog TSVKIT. Used as a manual delivery tool that engages the Fixture Mount Transfer (FMT) to transport the dental implant from the product packaging to the osteotomy.

Recall number
Z-1115-2015
Recalling firm
Zimmer Dental Inc
Product
GemLock Long Hex Driver, Catalog RHL2.5; Also distributed within the Tapered SwissPlus & SwissPlus Implant Systems Kits, Surgical, Complete, Catalog OPCST; Tapered Screw-Vent Implant System Surgical Kit, Complete, Catalog TSVKIT. Used as a manual delivery tool that engages the Fixture Mount Transfer (FMT) to transport the dental implant from the product packaging to the osteotomy.
Status
Terminated
Initiated
2015-01-22
Posted
2015-02-12
Terminated
2015-08-03
Reason
The Gemlock Long Hex Driver may not fit into the Fixture Mount Transfer (FMT) or the internal hex of the implant.
Root cause
Device Design
Action
An Urgent Device Recall Notice was sent on 1/22/15 to their domestic and foreign customers who purchased the Gemlock Long Hex Driver. The letter informs the customers of the problems identified, risks, and the actions to be taken. Customers are instructed to review the notice and check their inventory for the affected units. All affected product is to be immediately quarantined and removed from service. Customers are to contact Simmer Dental Customer Service at 1-800-854-7019 to obtain a Return Authorization Number and set up return of the products. Customers should complete the attached Business Reply Form and fax it to (574) 372-4265 or e-mail to corporatequality.postmarket@zimmer.com. Customers with questions are instructed to contact Zimmer Dental Customer Service at (800) 854-7019, Mon-Fri, 7 am-5pm (Pacific). Zimmer Dental is expanding the recall to include 1 additional lot for catalog no.TSVKIT. Zimmer Dental plans to inform an amended customer notification letter to their US and foreign consignees. The letter will go out to customers on 4/20/15 to inform them of the expanded lot.
Quantity
627 units
Distribution
Worldwide Distribution -- US, Argentina, Australia, Bulgaria, Canada, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Egypt, France, Germany, Hong Kong, Hungary, Iran, Israel, Italy, Japan, Lebanon, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Pakistan, Panama, Poland, Romania, Russia, Saudi Arabia, Serbia, Taiwan, Thailand, Tunisia, Turkey, and UAE. Expanded Distribution: Spain
Lot, serial or model codes
Etched Lot No. 62699693 & 62735177. Lot Number on Label: 62712431, 62743168, 62766610, 62702589, 62710484, 62712437, 62743075, 62750487, 62753787, 62756920, 62759862, 62760388, 62768456, and 62773158.  Expanded Lot No. 62728173.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NDP
Firm FEI
1038806
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register