FDA Medical Device Recalls (openFDA)
SunMed Holdings, LLC: Brand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Model/Catalog Number: 7700REA Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: No
- Recall number
- Z-1114-2026
- Recalling firm
- SunMed Holdings, LLC
- Product
- Brand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Model/Catalog Number: 7700REA Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: No
- Status
- Open, Classified
- Initiated
- 2025-12-15
- Posted
- 2026-01-28
- Terminated
- not on file
- Reason
- Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
- Root cause
- Device Design
- Action
- AirLife issued an URGENT: MEDICAL DEVICE RECALL NOTICE to its consignees on 12/15/2025 via email and US mail. The notice explained the issue, potential risk to the patient, and requested the following: "Customer immediate actions: Please take the following actions immediately: 1. Stop using the affected tapes o Immediately discontinue use of all AirLife Broselow Rainbow Tapes identified as 2025 Edition, 36-23446 Rev 3 Print Version. 2. Identify and segregate affected product o Examine your inventory and clinical areas for affected Broselow Rainbow Tapes (see Image B for identifying characteristics). o Remove all affected tapes from clinical service. o Segregate or quarantine the affected tapes to prevent further use. 3. Follow disposition instructions o Follow the instructions provided in the accompanying communication regarding discarding/destruction of the affected tapes. o Do not redistribute or place any affected tapes back into service. 4. Notify your internal users o Ensure that all clinicians and healthcare professionals within your organization who may use the Broselow Rainbow Tape are informed of this recall/field removal and understand that the affected Rev 3 tapes must not be used. 5. Notify downstream customers (if applicable) o If you have further distributed the affected tapes, please identify your customers/consignees and notify them of this recall/field removal promptly. o Your notification may be enhanced by including a copy of this letter. o If you have impacted product on hand, do not ship it; instead, hold it for discarding/destruction as instructed. 6. Complete and return the Response Form o Please complete and return the attached Response Form via e-mail to productquality@myairlife.com as soon as possible. Please complete and return Attachment B if you have affected product for discarding/destruction. This allows us to document your receipt of this recall notice and the status of affected product at
- Quantity
- 95300
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Singapore.
- Lot, serial or model codes
- Lot Code: Lot/Serial Number(s) 0004306540 0004316209 0004328895 0004329080 0004330876 0004331254 0004332994 0004338228 0004338229 0004338230 0004339717 UDI-DI Each-10889483588970 Case-30889483588974
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OKI
- Firm FEI
- 1314417
- Firm city
- Grand Rapids
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28