Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: syngo.via syngo.CT Cardiac Planning, Model Number 10496180 Product Usage: Syngo.CT Cardiac Planning and syngo CT. Cardiac Function are image analysis software packages for evaluating cardiac CT angiography (CTA) volume sets. The software packages are designed to support the physician in determining the functional parameters of the left and right ventricles, confirming the presence or absence of physicianidentified myocardial enhancement defects and evaluation documentation and follow-up of any such finding.

Recall number
Z-1113-2019
Recalling firm
Siemens Medical Solutions USA, Inc
Product
syngo.via syngo.CT Cardiac Planning, Model Number 10496180 Product Usage: Syngo.CT Cardiac Planning and syngo CT. Cardiac Function are image analysis software packages for evaluating cardiac CT angiography (CTA) volume sets. The software packages are designed to support the physician in determining the functional parameters of the left and right ventricles, confirming the presence or absence of physicianidentified myocardial enhancement defects and evaluation documentation and follow-up of any such finding.
Status
Terminated
Initiated
2019-03-21
Posted
not on file
Terminated
2022-01-11
Reason
There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.
Root cause
Software design
Action
Urgent Customer Safety Advisory Notice letters dated 3/20/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-888-7436.
Quantity
112 total
Distribution
US Nationwide Distribution in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, IA, IL, KS, KY, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
Lot, serial or model codes
Software versions VB20A or VB30A
510(k) numbers
["K170221"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register