Toshiba American Medical Systems Inc: TSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System; TXS-101/R Aquilion RXL Computed Tomography X-Ray System; TSX-301A Aquilion One Computed Tomography X-Ray System; TSX-301B Aquilion Premium Computed Tomography X-Ray System; TSX-302A Aquilion Prime Computed Tomography X-Ray System. This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head. The Aquilion RXL has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

Recall number
Z-1113-2014
Recalling firm
Toshiba American Medical Systems Inc
Product
TSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System; TXS-101/R Aquilion RXL Computed Tomography X-Ray System; TSX-301A Aquilion One Computed Tomography X-Ray System; TSX-301B Aquilion Premium Computed Tomography X-Ray System; TSX-302A Aquilion Prime Computed Tomography X-Ray System. This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head. The Aquilion RXL has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.
Status
Terminated
Initiated
2013-10-01
Posted
2014-03-03
Terminated
2015-03-03
Reason
Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the scan conditions used for contrast study may not start.
Root cause
Software design
Action
Toshiba sent an Urgent Medical Device Correction letter on October 1, 2013, via USPS return receipt mail and accompanied by a response form. The recall notification letter provided the Device Model, serial number, SID, problem, corrective action, request to customers, and contact information (800) 421-1968 or (800) 521-1968. Toshiba designed and test an upgraded ROM for the SC2 PWB to prevent recurrence of the fault. Service technicians will replace the ROM on site at no cost to the system owners. All repairs are expected to be completed by April 2014 and progress reports will be provided on the effectiveness of the plan. For questions regarding this recall call 714-730-5000.
Quantity
338 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
SERIAL NUMBER JDA1062176, JDA1062177, JDA1072183, JDA1072184, JDA1072186, JDA1072187, JDA1082188, JDA1082189, JDA1082190, JDA10X2191, JDA10X2192, JDA10X2193, JDA10X2194, JDA10X2195, JDA10X2196, JDA10X2197, JDA10X2198, JDA10Y2199, JDA10Y2200, JDA10Y2201, JDA10Y2202, JDA10Y2203, JDA10Y2204, JDA10Y2205, JDA10Z2206, JDA10Z2207, JDA10Z2209, JDA10Z2210, JDA10Z2211, JDA10Z2212, JDA10Z2213, JDA10Z2214, JDA10Z2215, JDA1112216, JDA1112217, JDA1112218, JDA1112219, JDA1112221, JDA1112222, JDA1112224, JDA1112225, JDA1112226, JDA1112227, JDA1122231, JDA1122234, JDA1142236, JDA1162237, JDA1162238, JDA1162239, JDA1182240, JDA1182241, JDA1182242, JDA1182243, JDA1182244, JDA11X2245, JDA11X2246, JDA11X2247, JDA11Z2249, JDA11Z2250, JDA11Z2252, JDA11Z2253, JDA11Z2254, JDA11Z2255, JDA1222257, JDA1222258, JDA1222259, JDA1222260, JDA1222261, JDA1242264, JDA1252265, JDA1252266, JDA1252267, JDA12Z2271, JDA1312272, JGA1062178, JGA1062181, JGA1062182, JGA1112220, JGA1112228, JGA11Z2256, JGA1222262, JGA12X2269, JLA1062179, JLA1062180, JLA1072185, JLA10Z2208, JLA1122232, JLA1122233, JLA1132235, JLA11Y2248, JLA11Z2251, JLA1232263, JLA1262268, JLA12Y2270, KDA1062080, KDA1072081, KDA1072082, KDA1072083, KDA1072085, KDA1072086, KDA1072088, KDA1082090, KDA1082091, KDA1082092, KDA1092093, KDA10X2094, KDA10X2095, KDA10X2096, KDA10X2097, KDA10X2098, KDA10X2099, KDA10Y2100, KDA10Y2101, KDA10Y2102, KDA10Z2104, KDA1112107, KDA1122110, KDA1122111, KDA1132113, KDA1142114, KDA1152115, KDA1152116, KDA1152117, KDA1152118, KDA1162119, KDA1162120, KDA1162121, KDA1162122, KDA1162123, KDA1162124, KDA1172125, KDA1182126, KDA1182127, KDA1182128, KDA1182129, KDA1192130, KDA1192131, KDA11X2132, KDA11X2133, KDA11Y2134, KDA11Z2136, KDA11Z2137, KDA11Z2138, KDA11Z2139, KDA11Z2140, KDA1222141, KDA1222142, KDA1222143, KDA1232144, KDA1242145, KDA1242146, KDA1252147, KDA1272148, KDA1282149, KDA1292151, KDA12X2152, KDA12Y2153, KDA12Z2154, KDA1322155, KGA1082089, KGA10Z2103, KGA10Z2105, KGA1112106, KGA1112108, KGA1122109, KGA1122112, KGA11Z2135, KLA1072084, KLA1072087, KLA1292150, RDA1292019, RDA12X2023, RDA12X2024, RDB12Y2029, RDB12Z2032, RDB1312037, RDB1312038, RGA12X2025, RGB12Y2028, RGB1312034, 2DA0792002, 2DA0792004, 2DA07X2005, 2DA07X2008, 2DA07Y2010, 2DA0842023, 2DA0842026, 2DA0862034, 2DA0862036, 2DA0872038, 2DA0882048, 2DA0882049, 2DA0882051, 2DA0882052, 2DA0882053, 2DA08X2061, 2DA08X2067, 2DA08X2070, 2DA08X2071, 2DA08X2072, 2DA08Y2077, 2DA08Y2080, 2DA08Z2092, 2DA0952115, 2DA0962116, 2DA0972123, 2DA0982131, 2DA0982133, 2DA09X2139, 2DA09X2143, 2DA09Y2148, 2DA09Y2151, 2DA09Z2158, 2DA09Z2161, 2DA09Z2162, 2DA1012165, 2DA1012167, 2DA1012168, 2DA1012169, 2DA1012170, 2DA1012174, 2DA1062208, 2DA1072212, 2DB1082219, 2DB1082220, 2DB1082222, 2DB10X2241, 2DB10Y2253, 2DB10Y2257, 2DB10Z2265, 2DB1112275, 2DB1112283, 2DB1122284, 2DB1122291, 2DB1122294, 2DC1142308, 2DC1142311, 2DC1152316, 2DC1162324, 2DC1172328, 2DC1172332, 2DC1172333, 2DC11X2352, 2DC11Y2365, 2DC11Y2367, 2DC11Z2372, 2DC11Z2374, 2DC1212377, 2DC1212383, 2DC1232392, 2DC1232393, 2DC1232396, 2DC1232397, 2DC1232398, 2DC1242407, 2DC1242409, 2DC1252410, 2DC1262412, 2DC1272417, 2DC1272421, 2DC1282423, 2DC12X2427, 2DC12Y2430, 2DC12Y2431, 2DC12Z2433, 2DC1312435, 2DC1312436, 2DC1322437, 1DA0962006, 1DA0972007, 1DA0972008, 1DA0992011, 1DA09Y2014, 1DA1062015, 1DB1072016, 1DB1082017, 1DB1082018, 1DB1082019, 1DB10Y2020, 1DB10Y2021, 1DB10Y2022, 1DB1112025, 1DB1122027, 1DC1142030, 1DC1152031, 1DC1162032, 1DC1162033, 1DC1172035, 1DC11Y2037, 1DC11Y2038, 1DC11Z2040, 1DC1222042, 1DC1232043, 1DC1232044, 1DC1252045, 1DC1252046, 1DC1262048, 1DC1272050, 1DC1282051, 1DC1292052, 1DC12X2053, 1DC12Y2054, 1DC12Y2055, 1DC12Z2056, 1DC1312057, 1DC1312058, 1DC1342059, 1GC1262047, 1GC1272049, 1DA1222096, 1DA1222097, 1DA1232099, 1DA1232100, 1DA1232103, 1DA1252124, 1DA1252131, 2DA1252001, 2DA1262003, 2DA1272004, 2DA1272005, 2DA1272006, 2DA1282007, 2DA1292008, 2DA12X2009, 2DA12X2010, 2DA12X2012, 2DA12X2013, 2DA12Y2014, 2DA12Y2015, 2DA12Y2016, 2DA12Z2017, 2DA12Z2018, 2DA1312019, 2DA1312020, 2DA1312021, 2DA1312022, 2GA1262002, 2GA12X2011, 2DA1072211, 2DA08X2063, 2DC11Y2368.
510(k) numbers
["K080211","K113466","K121553"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register